• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER NEUROVASCULAR CORK SURPASS EVOLVE 4.0MM X 20MM - PMAS; INTRACRANIAL ANEURYSM FLOW DIVERTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER NEUROVASCULAR CORK SURPASS EVOLVE 4.0MM X 20MM - PMAS; INTRACRANIAL ANEURYSM FLOW DIVERTER Back to Search Results
Model Number FDS40020
Medical Device Problem Code Activation Failure (3270)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/10/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during the procedure the subject stent deployed half in catheter and half in vessel.The device was removed from the patient and replaced.The procedure was completed successfully with no clinical consequences to the patient.
 
Additional Manufacturer Narrative
Device discarded at facility.
 
Additional Manufacturer Narrative
H4 manufacturing date ¿ added d4 expiration date - added due to the automated mes/dhr system there are controls in the manufacturing process to ensure the product met specifications upon release.The subject device is not available; therefore, visual and functional testing as well as physical analysis cannot be performed.The reported complaint could not be confirmed, and it could not be definitively determined if the device failed to meet specifications because the product was not returned.The reported event is covered in the device directions for use (dfu).As well, the risk of the reported event is documented in the risk documentation and there are current controls to mitigate the risk of the as reported event.Additional information provided by the customer indicated the device was confirmed to be in good condition prior to use and preparation/set-up was performed as per the directions for use.Continuous flush was maintained for the duration of the procedure.The stent was not fully deployed out of the catheter and had partial deployment (half in catheter, half in vessel).There was no resistance encountered during the procedure.No excessive force was applied at any point while advancing the stent within the microcatheter.The procedure was completed by opening another device of the same diameter and length.While there are a number of potential causes for the reported issue, because review and analysis of available information failed to identify a definitive cause, a cause of undeterminable was assigned for the reported event of the stent partial deployment.Device not returned for evaluation.
 
Event or Problem Description
It was reported that during the procedure the subject stent deployed half in catheter and half in vessel.The device was removed from the patient and replaced.The procedure was completed successfully with no clinical consequences to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SURPASS EVOLVE 4.0MM X 20MM - PMAS
Common Device Name
INTRACRANIAL ANEURYSM FLOW DIVERTER
Manufacturer (Section D)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI  NA
Manufacturer (Section G)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI   NA
Manufacturer Contact
tara lopez
47900 bayside parkway
fremont, CA 94538
5104132500
MDR Report Key13728882
Report Number3008881809-2022-00114
Device Sequence Number20035495
Product Code OUT
UDI-Device Identifier07613327375206
UDI-Public07613327375206
Combination Product (Y/N)N
Initial Reporter StateSC
Initial Reporter CountryUS
PMA/510(K) Number
P170024/S003
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/18/2024
Device Model NumberFDS40020
Device Catalogue NumberFDS40020
Device Lot Number22748426
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/10/2022
Supplement Date Received by Manufacturer05/18/2022
Initial Report FDA Received Date03/11/2022
Supplement Report FDA Received Date06/16/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/19/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-