Catalog Number 8606500 |
Device Problems
Gas Output Problem (1266); Failure to Deliver (2338); Intermittent Communication Failure (4038)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/11/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
The investigation has just begun; results will be provided in a follow-up report.
|
|
Event Description
|
It was reported that a ventilator failure occurred during a surgical procedure.Reportedly, this did not lead to any consequences for the patient.
|
|
Event Description
|
It was reported that a ventilator failure occurred during a surgical procedure.Reportedly, this did not lead to any consequences for the patient.
|
|
Manufacturer Narrative
|
The log file analysis revealed that during the case in question the device detected an internal communication error onboard a pcb that controls the communication between user interface, gas mixer and ventilator.The workstation initiated an emergency shutdown of automatic ventilation and posted corresponding alarms.Monitoring and manual ventilation remain possible in this state.The general issue is known from earlier occurrences of the same phenomenon.Intensive evaluation of the provided information and testing of concerned materials did not result in any finding.The fact that the involved pcb is used in another functional unit of the entire device without showing similar symptoms makes a design problem rather unlikely.A reasonable explanation would be emc disturbances beyond the immunity barriers of the workstation which is compliant to iec 60601-1-2.Therefore it is recommended that the conditions at the user facility should be checked to prevent from emc disturbance.The field failure rate of the affected pcb is accepted.
|
|
Search Alerts/Recalls
|