| Model Number |
M0068403910 |
| Medical Device Problem Codes |
Break (1069); Degraded (1153)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/05/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The event was reported by a distributor, however the healthcare facility where the event occurred is as follows: (b)(6).(b)(4) investigation results: the returned flexiva laser fiber was analyzed, and a visual evaluation noted that the laser fiber was returned in two pieces.The first piece measured 16 cm and the second piece measured 293.7 cm.The exposed glass tip measured 3 mm and appeared in good condition.A functional evaluation revealed that there was no light from the sma connector, indicating a fracture within the connector.No other problems with the device were noted.The reported event of the flexiva laser fiber failed to emit energy was confirmed.The reported failure to emit laser energy is consistent with a fiber broken within the connector.It is likely that procedural handling of the device during set-up or use contributed to the fracture within the fiber connector.A broken fiber within the connector of the device can result in insufficient aiming beam and low laser energy output, up to a complete inability for the fiber to fire.Review and analysis of all available information indicated that the most probable root cause is adverse event related to procedure.A review of the manufacturing documentation for this device revealed that no anomalies or deviations related to the event occurred during manufacturing.A labeling review was performed and, from the information available, this device was used per the instructions for use (ifu) / product label.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation on october 14, 2021 that a flexiva laser fiber was used during a procedure , performed on (b)(6) 2021.During the procedure, the fiber stopped working three times and did not work anymore.The procedure was completed with another flexiva laser fiber.There were no patient complications reported as a result of this event.Boston scientific has been unable to obtain additional information regarding the event to date, despite good faith efforts.Note: this event has been deemed reportable based on the investigation finding that the laser fiber was broken within the connector and fiber body was broken.
|
| |
|
Search Alerts/Recalls
|