• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. TPRLC 133 T1 PPS SO 10X140MM T1; PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. TPRLC 133 T1 PPS SO 10X140MM T1; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Medical Device Problem Codes Device Contaminated During Manufacture or Shipping (2969); Manufacturing, Packaging or Shipping Problem (2975)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/18/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during inspection of circulated products in the warehouse, the product was identified as having damaged sterile packaging.Upon initial review of the pictures provided, a hair like substance was present in the packaging.No patient involvement and no adverse consequences reported.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Additional Manufacturer Narrative
(b)(4).Report source: foreign: (b)(6).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Visual evaluation of the returned product identified damage to the sterile packaging (blister).Sterility has not been compromised.Evaluation of the returned product confirmed there is white debris inside the sterile packaging which is consistent with the appearance of foam debris from the foam packaging inside the sterile barrier.Device history record was reviewed and no discrepancies relevant to the reported event were found.The likely condition of the device when it left zimmer biomet is conforming to specification.The root cause of the damaged packaging can be attributed to transit damage, and a packaging design issue.The root cause of the debris in sterile packaging is likely to be due to transit damage causing the foam packaging to become abraded and shed, and a packaging design issue.A corrective action was previously opened to address the reported event.This event is no longer considered reportable.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TPRLC 133 T1 PPS SO 10X140MM T1
Common Device Name
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key13731738
Report Number0001825034-2022-00579
Device Sequence Number14608439
Product Code KWA
UDI-Device Identifier00880304489684
UDI-Public(01)00880304489684(17)300220(10)6743454
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K101086
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 05/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number51-103100
Device Lot Number6743454
Was Device Available for Evaluation? Device Returned to Manufacturer
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/18/2022
Supplement Date Received by Manufacturer04/29/2022
Initial Report FDA Received Date03/11/2022
Supplement Report FDA Received Date05/04/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/20/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Patient SexUnknown
-
-