(b)(4).Concomitant medical products: cat#00801803601 lot#63474815.Cat# 00620005220 lot# 693315330.Cat# 00811400100 lot# 63885445.Report source: foreign: country: (b)(6).Reported event was confirmed due to the review of medical records.Radiographs were provided and reviewed by a health care professional.Review of the available records identified the following: left total hip arthroplasty with undersized acetabular cup.Third image demonstrate superior dislocation.It was mentioned by the hcp reviewing the x-rays that the shell appeared to be undersized and likely led to the dislocation, but this cannot be confirmed.Device history record (dhr) was reviewed and no discrepancies were found.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.Multiple mdr reports were filed for this event, please see associated reports: 0002648920 - 2021 - 00451.
|