Model Number N/A
Medical Device Problem Codes Loosening of Implant Not Related to Bone-Ingrowth (4002); Migration (4003)
Health Effect - Clinical Code Pain (1994)
Date of Event 02/22/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Concomitant medical products: item #: unknown, unknown head, lot #: unknown.Item #: unknown, unknown stem, lot #: unknown.Item #: unknown, unknown liner, lot #: unknown.Report source: (b)(6).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported that the patient underwent a hip revision approximately five months post implantation due to pain, loosening of implant, and implant migration.Optimal positioning and augmentation with screw can be seen on immediate post op film.Follow up x-rays at six weeks post op showed changed in cup position secondary to failure of ingrowth and loosening.Patient started experiencing increasing hip pain at five weeks post op.Revision was finally done after a trial of conservative management with protected weight bearing did not help ingrowth of cup.Patient continued to experience worsening pain.Cup was revised.No additional information is available.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.A shell and a screw were returned for evaluation.Shell was returned with liner fully seated and fractured screw head still in the shell hole.Screw head was not removed for evaluation.Shell shows nicks, gouges, and scratches to the outer diameter of the face.Bio debris discoloration is noted to the fiber metal surface.Visual review of the fractured bone screw shows screw has fractured.(2) fractured pieces were returned.Proximal screw head remains in the shell hole between the liner and shell.Distal screw threads shows fractured surface and extraction damage including nicks, gouges, and deformation.The screw was sent to sem for further analysis.The analysis identified bone screw sample could not be determined due to the lack of fracture artifacts.Elemental analysis of the bone screw sample showed that it is consistent with ti-6al-4v alloy.Radiographs were provided and reviewed by a health care professional.Review of the available records identified incomplete integration of the acetabular cup with the subjacent bone with subsidence of the acetabular cup and likely acetabular cup screw breakage.Device history record was reviewed and no discrepancies relevant to the reported event were found.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
52MM O.D. SIZE II POROUS UNCEMENTED WITH MULTI-HOLES SHELL
Common Device Name
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key13731922
Report Number0001822565-2022-00702
Device Sequence Number17299150
Product Code LPH
UDI-Device Identifier00889024150188
UDI-Public(01)00889024150188(17)301031(10)64869971
Combination Product (Y/N)N
Initial Reporter CountrySN
PMA/510(K) Number
K151448
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number00875305202
Device Lot Number64869971
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/28/2022
Supplement Date Received by Manufacturer07/21/2022
Initial Report FDA Received Date03/11/2022
Supplement Report FDA Received Date07/25/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/06/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H10.
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexFemale
-
-