| Model Number |
03.404.018S |
| Medical Device Problem Code |
Entrapment of Device (1212)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
02/24/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Device report from synthes reports an event in (b)(6) as follows: it was reported that on (b)(6) 2022, during a scheduled transplant operation bone marrow, the practitioner used an ria kit 2 (boring, irrigation, suction) in frame masklet's technique.During the procedure the fins of the heads, bore sizes of 10, 10.5, 11 and 14 mm, broke in the distal third of the femur despite use of the device in accordance with supplier's recommendations.The 16 broken fins ended up in intraosseous.The practitioner was able to remove about half of the fins of the boring heads; however, parts remain in the femur despite numerous recovery attempts.There was a surgical delay of forty-five (45) minutes.The procedure was successfully completed.This report involves one (1) reamer.This is report 2 of 5 for (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.Reporter is a j&j employee.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Investigation summary: the product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Additionally, a visual inspection of the returned device was performed from attachment (image1.Jpg).Visual analysis of the returned sample revealed that the 11.0mm reamer head for ria 2 sterile was found to be broken from the shaft.Embedded device condition cannot be confirmed since x-ray evidence was not provided.A dimensional inspection for the 11.0mm reamer head for ria 2 sterile was unable to be performed due to post manufacturing damage.The observed condition of the device was consistent with a random component failure that may have been caused by exposure to unintended forces.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the observed condition of the 11.0mm reamer head for ria 2 sterile would contribute to the complained device issue.Due to the trend with the broken ria 2 reamer heads identified during post-market surveillance, ae capa was raised to determine the root cause of broken head breakages.The cause of this issue was identified as deviation from the recommended surgical approach of 10 degrees.Additionally, the product issue escalation was initiated to define any further action related to the breakage.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: production order traveler met all inspection acceptance criteria.Packaging label log was reviewed and determined to be conforming.Packaging bom was reviewed and determined to be conforming with no deviations to normal packaging identified.This lot met all dimensional, visual, sterility and packaging criteria at the time of release with no issues documented during the inspection or release of the product that would contribute to this complaint condition.Component part(s) reviewed: device history review: this lot met all dimensional, visual, sterility and packaging criteria with no issues documented during the inspection or release of the product that would contribute to this complaint condition.
|
| |
|
Search Alerts/Recalls
|
|