| Model Number |
01-EC-459-1401 |
| Medical Device Problem Code |
Gas/Air Leak (2946)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/17/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The actual complaint product is reported to be unavailable for evaluation.A review of the device history record is not possible as no lot number was provided.All information reasonably known as of 10 mar 2022 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical inc.Avanos medical inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical complaint database and identified as complaint (b)(4).
|
| |
|
Event or Problem Description
|
|
Avanos medical inc.Received a single report that referenced two different incidences, which were associated with separate units, involving the same patient.This is the first of two reports.Refer to 3011270181-2022-00017 for the second report it was reported that the device manifold was "leaky" and was replaced.There was no injury to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
The product involved in the report has been returned and the investigation remains in progress at this time.All information reasonably known as of 21 apr 2022 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical inc.Avanos medical inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical inc.Product is defective or caused serious injury.
|
| |
|
Additional Manufacturer Narrative
|
|
The actual sample from the reported event was returned for evaluation.Visual examination of the device revealed the sleeve was free of tearing or damages, the manifold was free of cracks, or any deformation and the irrigation assembly appeared intact; additionally, the control valve functioned as intended: the male and female swivels appeared intact and swiveled accordingly.Testing was completed by research and development (r&d) and technical quality, the performed test(s) confirmed there was leaking in the manifold.The reported event could be confirmed as reported; the root cause could not be determined.All information reasonably known as of 22 aug 2022 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical inc.Avanos medical inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical inc.Product is defective or caused serious injury.
|
| |
|