| Model Number |
786414 |
| Medical Device Problem Code |
Material Fragmentation (1261)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/15/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The device was not returned.
|
| |
|
Event or Problem Description
|
|
It was reported that the ureteral stent underwent catheter breakage when used by the clinician.A new stent was used.
|
| |
|
Event or Problem Description
|
|
It was reported that the ureteral stent underwent catheter breakage when used by the clinician.A new stent was used.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event is inconclusive because no sample was returned for evaluation.A potential root cause for this event could be, "material selection".The device was used for treatment purposes, however, it is unknown if the device had met relevant specifications or contributed to the reported event.No manufacturing issues or non-conformance's were noted in the dhr review that could have caused or contributed to the reported event.Based on the results of the investigation no additional action is required at this time.The device was not returned for evaluation.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "avoid improper handling of stent such as bending, kinking, tearing, etc.Misuse could damage the overall integrity of the stent." "care should be exercised when removing the stent from the inner polybag to eliminate tearing or fragmentation." "the insertion of a ureteral stent should only be done by those individuals who have comprehensive training in the techniques and risks of the procedure." h11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the device was not returned.
|
| |
|
Search Alerts/Recalls
|