Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.Reporter is a j&j employee.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Device report from synthes reports an event in taiwan as follows: it was reported that on (b)(6) 2022, surgeon used an surgical instrument (screw tap) to form a thread on the patient's bone, the patient's bone was too hard.The front end of tap was broken into the patient's l5 vertebral body.The surgeon could not remove it after repeated attempts and it remained in patient's body.This report is for one (1) viper2 5mm self drilling tap.This is report 1 of 1 for (b)(4).
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: h3, h6: a review of the receiving inspection (ri) for viper2 5mm self drilling tap, was conducted identifying that lot number ng78962 was released in a four batch.Part: 286715500, lot: ng78962, supplier: ng instruments.Batch1: released on november 19, 2018 with no discrepancies.The ri identified no issues during the manufacturing and release of this device that could have contributed to the problem reported by the customer.Visual analysis of the returned sample revealed that the tip of the viper2 5mm self drilling tap has broken off.No other product defect identified.No evidence indicating the broken fragment remains inside patient was provided.A dimensional inspection was performed for the viper2 5mm self drilling tap, and met specifications.The observed condition of the device was consistent with a random component failure that may have been caused by exposure to unintended forces as mentioned in the event description.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the observed condition of the viper2 5mm self drilling tap, would contribute to the complained device issue.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Drawing/specifications reviewed current and manufactured revisions were reviewed.Dimensional inspection: measured dimension: conforming.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|