• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 SAFESTEP HUBER NEEDLE SET 22G X 1 IN WITH Y-SITE; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 SAFESTEP HUBER NEEDLE SET 22G X 1 IN WITH Y-SITE; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Model Number N/A
Medical Device Problem Codes Fluid/Blood Leak (1250); Defective Component (2292)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/27/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The manufacturer has received the sample and will evaluate.Results are expected soon.A lot history review (lhr) of asevf018 showed one other similar product complaint(s) from this lot number.The complaints for this lot number have been reported from the same facility in (b)(6).
 
Event or Problem Description
It was reported that when the safestep was accessed to the port body and blood return was confirmed, air contamination was found.Therefore, the device was removed.Another needle was used to confirm for blood return after accessing to the port body again, but air contamination was found again.There was no reported patient injury.This report addresses the first device used.
 
Event or Problem Description
It was reported that when the safestep was accessed to the port body and blood return was confirmed, air contamination was found.Therefore, the device was removed.Another needle was used to confirm for blood return after accessing to the port body again, but air contamination was found again.There was no reported patient injury.This report addresses the first device used.
 
Additional Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The following were reviewed as part of this investigation: patient severity, complaint and lot history, applicable previous investigation(s), labeling, applicable manufacturing records, and applicable fmea documents.Based on a review of this information, the following was concluded: the complaint that air was able to enter the infusion set was confirmed and the damage appeared to be supplier related.Two 22g safestep infusion sets with y-injection site was returned for investigation.A functional test revealed that air could enter one infusion set through the y-site when a negative pressure was applied to the infusion set.No continuous leak of fluid was observed in any part of the returned samples when the infusion sets were pressurized with water.The affected valve was cut open and the valve stem was removed, which revealed flow lines in the sealing surface of the valve stem and pitting in other areas of the valve stem.The damage to the sealing surface of the valve stem was located in an area that was inaccessible to the user.The uneven areas of the valve stem sealing surface most likely allowed air to bypass the valve during aspiration.Bd is working closely with the supplier to prevent recurrence of the reported event.H3 other text : evaluation findings are in section h.11.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SAFESTEP HUBER NEEDLE SET 22G X 1 IN WITH Y-SITE
Common Device Name
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer Contact
johanna de oliveira
605 north 5600 west
salt lake city 84116
8015950700
MDR Report Key13736974
Report Number3006260740-2022-00757
Device Sequence Number5069495
Product Code FPA
UDI-Device Identifier00801741066191
UDI-Public(01)00801741066191
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K153440
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue NumberLH-0030YN
Device Lot NumberASEVF018
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/07/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/16/2022
Supplement Date Received by Manufacturer04/27/2022
Initial Report FDA Received Date03/11/2022
Supplement Report FDA Received Date05/07/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/01/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-