| Model Number |
P8.5-80D245 |
| Medical Device Problem Code |
Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Information was received that the patient underwent a revision procedure due to the rod/magnet failing.No patient adverse event was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The device has not been returned for evaluation.The root cause is unable to be determined at this time.If any additional information is provided, a supplemental report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported defect code was confirmed as the visual inspection of the received x-rays revealed that the thrust bearing broke away from the gear stage which caused the nail to stop lengthening.Device records review: review of the device history records (dhr) indicated the unit met all quality specifications and testing prior to product departure.
|
| |
|
Event or Problem Description
|
|
See h10.
|
| |
|