Manufacturer¿s ref.No: (b)(4).Information regarding patient identifier, date of birth, age, gender, weight, race, and ethnicity were not provided.The initial reporter phone: (b)(6).The initial reporter email address is not available / reported.Conclusion: the healthcare professional reported that a patient underwent a mechanical thrombectomy of a middle cerebral artery (mca) m2 segment occlusion with a 5mm x 33mm embotrap ii revascularization device (et009533 / 21g174av) and experienced bleeding after the embotrap ii device was retrieved.The complaint device is not available to be returned.A review of the manufacturing documentation associated with this lot (21g174av) top and sub assembly presented no issues during the manufacturing or inspection processes related to the reported complaint.There were no non-conformances related to device manufacture or inspection.All product rejected during manufacturing was identified as scrap and properly accounted for.No determination of causes and possible contributing factors could be made.As such, the investigation will be closed.Hemorrhage secondary to vascular injury is a well-known potential adverse event associated with endovascular mechanical thrombectomy for acute ischemic stroke cases and is listed in the embotrap ii instructions for use (ifu) as such.With the information provided, it is not possible to determine the root cause of the event.However, there are vessel characteristics, patient and procedural factors that may have contributed to the reported event.There is no indication that the device malfunctioned, or that the event is related to a device design or manufacturing issue.Intracranial hemorrhage is a condition which can cause permanent damage/death or may require additional intervention to prevent further/permanent injury.Furthermore, the relationship of the embotrap ii to the reported event cannot be excluded.Based on the review of the manufacturing documentation, there is no indication that the event is related to the device manufacturing process.As part of the post market surveillance program, information from this complaint is trended for statistical signals and corrective / preventive action may be triggered at a later time.Since there was no evidence to suggest the event was related to a manufacturing or design issue, no corrective actions will be taken at this time.The manufacturer will submit a supplemental report if new facts arise which materially alter information submitted in a previous mdr report.Additional information will be submitted within 30 days of receipt.
|
|
Manufacturer¿s ref.No: (b)(4).The purpose of this mdr submission is to include the additional event information received on 16-mar-2022 that changed the reportability of the event as related to this product reported in this medwatch report.[additional information]: on 16-mar-2022, additional information was received.The information indicated that the physician ¿mistook this case to another case which a competitor¿s device was used.This case had no adverse event.There was no bleeding occurred in this case.¿ upon further review, this complaint does not meet the criteria for medical device reporting since there was no adverse event and there is no complaint against a cerenovus device.This event is no longer considered us fda mdr reportable.No further reports will be forthcoming.
|