• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 GROSHONG NXT CLEARVUE 4F SINGLE-LUMEN PICC BASIC TRAY; CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 GROSHONG NXT CLEARVUE 4F SINGLE-LUMEN PICC BASIC TRAY; CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS Back to Search Results
Model Number N/A
Device Problem Fluid/Blood Leak (1250)
Patient Problems Infiltration into Tissue (1931); Fluid Discharge (2686); Swelling/ Edema (4577)
Event Date 02/22/2022
Event Type  malfunction  
Event Description
It was reported that this patient, was diagnosed with rhinocarcinoma.On (b)(6) 2021, a groshong catheter was placed in the right chelidon for the first time, with 39 cm placed internally and another 5 cm exposed externally.The puncture was successful.When infusing immune stimulants at outpatient on (b)(6) 2022, exudation and swelling were observed.The nurse for catheter maintenance gently pulled the catheter and found a large damage (nearly rupture) with the catheter at the scale of approximately 2-4 cm.The catheter was removed eventually due to the serious damage.Additional information: no treatment necessary.
 
Manufacturer Narrative
The device has not been returned to the manufacturer for evaluation.A lot history review (lhr) of reen1971 showed no other similar product complaint(s) from this lot number.
 
Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The following were reviewed as part of this investigation: patient severity, applicable previous investigation(s), returned sample, applicable fmea documents, and applicable manufacture records.Based on a review of this information, the following was concluded: the complaint of a leaking catheter was confirmed.The product returned for evaluation was 4fr s/l groshong.The investigation findings are consistent with damage accumulated through flexural fatigue.Flexural fatigue occurs due to cyclic kinking of the catheter tube in which physiological, placement, usage, and mechanical factors may gradually form a crack(s) in the catheter.The returned product sample was evaluated and a split was observed at the 37cm depth marking.The catheter split contained physical features associated with material fatigue, and the characteristics observed which supported this type of failure included: ¿ damage which was circumferentially aligned ¿ fracture edges which were rounded and polished due to repeated material wear ¿ 'c' shaped impressions leading into the fracture site which are consistent with material buckling due to movement which caused the fracture edges to be pressed together ¿ overall elliptical shape to the fracture cross-section (a result of repeated kinking of the tubing) an examination of the catheter structure revealed no potential damage/defect related to manufacture of the product.The damage location suggested that catheter securement, access and maintenance techniques may have contributed.H3 other text : evaluation findings are in section h.11.
 
Event Description
T was reported that this patient, was diagnosed with rhinocarcinoma.On (b)(6) 2021, a groshong catheter was placed in the right chelidon for the first time, with 39 cm placed internally and another 5 cm exposed externally.The puncture was successful.When infusing immune stimulants at outpatient on (b)(6) 2022, exudation and swelling were observed.The nurse for catheter maintenance gently pulled the catheter and found a large damage (nearly rupture) with the catheter at the scale of approximately 2-4 cm.The catheter was removed eventually due to the serious damage.Additional information: no treatment necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GROSHONG NXT CLEARVUE 4F SINGLE-LUMEN PICC BASIC TRAY
Type of Device
CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
kelsey erickson
605 north 5600 west
salt lake city 84116
8015950700
MDR Report Key13737595
MDR Text Key287128285
Report Number3006260740-2022-00762
Device Sequence Number1
Product Code LJS
UDI-Device Identifier00801741035272
UDI-Public(01)00801741035272
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K034020
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 08/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/11/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number7655405
Device Lot NumberREEN1971
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/16/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/01/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age62 YR
-
-