C.R. BARD, INC. (BASD) -3006260740 GROSHONG NXT CLEARVUE 4F SINGLE-LUMEN PICC BASIC TRAY; CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS
|
Back to Search Results |
|
Model Number N/A |
Device Problem
Fluid/Blood Leak (1250)
|
Patient Problems
Infiltration into Tissue (1931); Fluid Discharge (2686); Swelling/ Edema (4577)
|
Event Date 02/22/2022 |
Event Type
malfunction
|
Event Description
|
It was reported that this patient, was diagnosed with rhinocarcinoma.On (b)(6) 2021, a groshong catheter was placed in the right chelidon for the first time, with 39 cm placed internally and another 5 cm exposed externally.The puncture was successful.When infusing immune stimulants at outpatient on (b)(6) 2022, exudation and swelling were observed.The nurse for catheter maintenance gently pulled the catheter and found a large damage (nearly rupture) with the catheter at the scale of approximately 2-4 cm.The catheter was removed eventually due to the serious damage.Additional information: no treatment necessary.
|
|
Manufacturer Narrative
|
The device has not been returned to the manufacturer for evaluation.A lot history review (lhr) of reen1971 showed no other similar product complaint(s) from this lot number.
|
|
Manufacturer Narrative
|
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The following were reviewed as part of this investigation: patient severity, applicable previous investigation(s), returned sample, applicable fmea documents, and applicable manufacture records.Based on a review of this information, the following was concluded: the complaint of a leaking catheter was confirmed.The product returned for evaluation was 4fr s/l groshong.The investigation findings are consistent with damage accumulated through flexural fatigue.Flexural fatigue occurs due to cyclic kinking of the catheter tube in which physiological, placement, usage, and mechanical factors may gradually form a crack(s) in the catheter.The returned product sample was evaluated and a split was observed at the 37cm depth marking.The catheter split contained physical features associated with material fatigue, and the characteristics observed which supported this type of failure included: ¿ damage which was circumferentially aligned ¿ fracture edges which were rounded and polished due to repeated material wear ¿ 'c' shaped impressions leading into the fracture site which are consistent with material buckling due to movement which caused the fracture edges to be pressed together ¿ overall elliptical shape to the fracture cross-section (a result of repeated kinking of the tubing) an examination of the catheter structure revealed no potential damage/defect related to manufacture of the product.The damage location suggested that catheter securement, access and maintenance techniques may have contributed.H3 other text : evaluation findings are in section h.11.
|
|
Event Description
|
T was reported that this patient, was diagnosed with rhinocarcinoma.On (b)(6) 2021, a groshong catheter was placed in the right chelidon for the first time, with 39 cm placed internally and another 5 cm exposed externally.The puncture was successful.When infusing immune stimulants at outpatient on (b)(6) 2022, exudation and swelling were observed.The nurse for catheter maintenance gently pulled the catheter and found a large damage (nearly rupture) with the catheter at the scale of approximately 2-4 cm.The catheter was removed eventually due to the serious damage.Additional information: no treatment necessary.
|
|