| Model Number |
IPN043837 |
| Medical Device Problem Code |
Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/15/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that "the material [tears]".No patient involvement reported.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Complaint verification testing could not be performed as the customer has reported that the sample is not available for investigation; however, the customer did provide photos.The photos were reviewed and it was confirmed that the tubing was torn.A device history record review was performed and no relevant findings were identified.The manufacturing site reports: based on the observation, this complaint is confirmed.The defect part was supplied by our supplier.In manufacturing, a non-conformance (nc) has been raised to address this issue and root causes with corrective actions will be implemented through this nc process.
|
| |
|
Event or Problem Description
|
|
It was reported that "the material [tears]".No patient involvement reported.
|
| |
|