Model Number IPN043837
Medical Device Problem Code Material Split, Cut or Torn (4008)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/15/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that "the material [tears]".No patient involvement reported.
 
Additional Manufacturer Narrative
(b)(4).
 
Additional Manufacturer Narrative
(b)(4).Complaint verification testing could not be performed as the customer has reported that the sample is not available for investigation; however, the customer did provide photos.The photos were reviewed and it was confirmed that the tubing was torn.A device history record review was performed and no relevant findings were identified.The manufacturing site reports: based on the observation, this complaint is confirmed.The defect part was supplied by our supplier.In manufacturing, a non-conformance (nc) has been raised to address this issue and root causes with corrective actions will be implemented through this nc process.
 
Event or Problem Description
It was reported that "the material [tears]".No patient involvement reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HUDSON HUMID-VENT FLEX TUBE
Common Device Name
SET, TUBING AND SUPPORT, VENTI
Manufacturer (Section D)
TELEFLEX MEDICAL SDN. BHD.
perak, west malaysia
Manufacturer (Section G)
TELEFLEX MEDICAL SDN. BHD.
lot no : pt2577 jalen perusahaan
4 kamunting industrial estate
perak, west malaysia 34600
MY   34600
Manufacturer Contact
katharine tarpley
3015 carrington mill blvd
morrisville, NC 27560
MDR Report Key13737684
Report Number8040412-2022-00050
Device Sequence Number4737311
Product Code BZO
UDI-Device Identifier04026704388769
UDI-Public04026704388769
Combination Product (Y/N)N
Initial Reporter CountryGM
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative,Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Expiration Date04/30/2026
Device Model NumberIPN043837
Device Catalogue Number20011
Device Lot NumberKMH21F0394
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/15/2022
Supplement Date Received by Manufacturer04/06/2022
Initial Report FDA Received Date03/11/2022
Supplement Report FDA Received Date04/06/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/25/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-