| Model Number |
SCPX146TH |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Deformity/ Disfigurement (2360)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Device image evaluation completed on (b)(6) 2022.Upon visual evaluation of the image provided in the complaint, the tissue expander was observed deflated.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.As part of mentor¿s quality process, all devices are manufactured, inspected, and released to approved specifications.The mentor failure analysis lab has received the device for evaluation.The analysis has begun but is not complete at this time.When the investigational analysis has been completed, a supplemental report will be submitted.Reason for device explant and/or reoperation: deflation.Manufacturer¿s reference number: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a female who underwent an unspecified breast surgery with mentor cpx4 plus smooth tall height 750cc breast implant experienced unknown- sided deflated post procedure.As a result, patient underwent replacement with cat#: (b)(4), sn#: (b)(4) on (b)(6) 2022.
|
| |
|
Additional Manufacturer Narrative
|
|
On 07-apr-2022, mentor completed the investigation on the suspect medical device.Device evaluation summary: according to the information received, the patient experienced deflation in the breast tissue expander.The product was returned to mentor for evaluation.Mentor conducted a visual inspection, microscopic examination, and shell thickness of the returned device.Visual analysis of the returned sample revealed that the tissue expander was found to have a tear on the union between shell and patch on the anterior view.Additionally the shell was observed to be blue.Microscopic examination was performed, and the cause of the tear could not be identified.Therefore, a thickness measurement was conducted on the shell at the tear, and the thickness was within manufacturing specifications.As the authorization form for examination was not received the product evaluation lab couldn¿t identify a conclusion due to the specific nature of this deflation.The evaluation was limited to non-destructive testing.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified.Although no conclusion could be reach on the cause of the reported event, the instructions for use contain the following caution: rupture can occur at any time after implantation but is more likely to occur the longer the tissue expander is implanted.The following may cause tissue expander to rupture: stressing the tissue expander during implantation and weakening it; folding or wrinkling of the tissue expander shell; excessive force to the chest (e.G.During closed capsulotomy); trauma; compression during mammographic imaging; and severe capsular contracture.Breast tissue expander may also simply wear out over time.As part of mentor¿s quality process all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post market surveillance.Manufacturer¿s reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation completed on april 14 , 2022: device photo evaluation completed on april 19 , 2022: upon visual evaluation of the image provided in the complaint, the tissue expander was observed deflated.Additionally, the shell was observed to be blue.As stated in the product insert data sheet, is contraindicated to place drugs or substances inside the tissue expanders other than sterile saline for injection since this could compromise the product integrity.As part of mentor¿s quality process, all devices are manufactured, inspected, and released to approved specifications.,according to the information received, the patient experienced deflation in the cpx4 plus smooth tall height 750cc breast tissue expander.The product was returned to mentor for evaluation.Mentor conducted a visual inspection, microscopic examination, and shell thickness of the returned device.Visual analysis of the returned sample revealed that the cpx4 plus smooth tall height 750cc tissue expander was found to have a tear on the union between shell and patch on the anterior view.Additionally, the shell was observed to be blue.A microscopic examination was performed, and the cause of the tear could not be identified.Therefore a thickness measurement was conducted on the shell at the tear, and the thickness was within manufacturing specifications.As the authorization form for examination was not received the product evaluation lab couldn´t identify a conclusion due to the specific nature of this deflation.The evaluation was limited to non-destructive testing.Although no conclusion could be reached on the cause of the reported event, the instructions for use contain the following caution: causes of deflation of tissue expander include but are not limited to the following events: crease fold failure, intraoperative or postoperative trauma, excessive stresses or manipulations as may occur during normal, daily routines including customary and purposeful trauma such as that which can occur during vigorous exercise, athletics, etc.As stated in the product insert data sheet, is contraindicated to place drugs or substances inside the tissue expanders other than sterile saline for injection since this could compromise the product integrity.As part of mentor¿s quality process, all devices are manufactured, inspected, and released to approved specifications.Manufacturer¿s reference number: (b)(4).
|
| |
|