BAYLIS MEDICAL COMPANY INC. VERSACROSS RF WIRE; CATHETER, SEPTOSTOMY Back to Search Results
Model Number VXW-35-180-P
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Perforation (2001)
Date of Event 02/14/2022
Type of Reportable Event Serious Injury
Event or Problem Description
A case of aortic perforation leading to a procedural abortion was reported during a cryoablation atrial fibrillation procedure, where the versacross rf wire was used to attempt transseptal puncture.Due to complex patient anatomy, the physician attempted a more superior transseptal approach, and while advancing the rf wire, the physician noticed an artifact and maneuvered an ice catheter to investigate.Upon investigation, the rf wire was observed to be in the aorta.The physician immediately requested a surgical consult, after which the rf wire was removed from the aorta without further complications.The procedure was aborted and the patient was admitted to the icu for overnight observation where they were later sent home.
 
Additional Manufacturer Narrative
A dhr review was completed for the lot in question, and it was confirmed that all devices met relevant requirements prior to release.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERSACROSS RF WIRE
Common Device Name
CATHETER, SEPTOSTOMY
Manufacturer (Section D)
BAYLIS MEDICAL COMPANY INC.
5959 trans-canada highway
montreal, quebec H4T 1 A1
CA  H4T 1A1
Manufacturer (Section G)
BAYLIS MEDICAL COMPANY INC.
2775 matheson blvd. east
mississauga, ontario L4W 4 P7
CA   L4W 4P7
Manufacturer Contact
meghal khakhar
2775 matheson blvd. east
mississauga, ontario L4W 4-P7
CA   L4W 4P7
MDR Report Key13739460
Report Number9710452-2022-00012
Device Sequence Number4762358
Product Code DXF
UDI-Device Identifier00685447003298
UDI-Public(01)00685447003298(17)240928(10)VXFH170921
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K150709
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/28/2024
Device Model NumberVXW-35-180-P
Device Lot NumberVXFH170921
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/14/2022
Supplement Date Received by Manufacturer02/14/2022
Initial Report FDA Received Date03/11/2022
Supplement Report FDA Received Date03/12/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured10/19/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization;
Patient SexUnknown
-
-