• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARROW INTERNATIONAL INC. AC3 OPTIMUS IABP NA/EMEA; SYSTEM, BALLOON, INTRA-AORTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARROW INTERNATIONAL INC. AC3 OPTIMUS IABP NA/EMEA; SYSTEM, BALLOON, INTRA-AORTIC Back to Search Results
Model Number IPN001112
Device Problem Power Problem (3010)
Patient Problem Insufficient Information (4580)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).There is an investigation ongoing into this reported complaint.A supplement follow-up mdr report will be submitted with the investigation results upon completion of the investigation.
 
Event Description
It was reported by the field service representative (fsr) that the pump shut off during use.
 
Event Description
It was reported by the field service representative (fsr) that the pump shut off during use.
 
Manufacturer Narrative
Qn# (b)(4).Teleflex received the device for investigation.The reported complaint of flashing power switch and 5 volt alarm was confirmed by the field service agent.Upon return, the usb a female connector was noted damaged, and a connector pin was noted bent, which could cause a short to ground.A definite root cause is undetermined; however, a potential cause is excessive force while inserting the mating usb in the wrong orientation.A device history record (dhr) review was conducted for the lot number with no relevant findings.The device passed all manufacturing specifications prior to release.Teleflex assessed the risk for the reported complaint.There are no new or revised risk.This will be monitored for any developing trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AC3 OPTIMUS IABP NA/EMEA
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC
Manufacturer (Section D)
ARROW INTERNATIONAL INC.
reading PA
Manufacturer (Section G)
ARROW INTERNATIONAL INC.
16 elizabeth drive
chelmsford MA 01824
Manufacturer Contact
kaylia pen
16 elizabeth drive
chelmsford, MA 01824
MDR Report Key13739678
MDR Text Key287347165
Report Number3010532612-2022-00052
Device Sequence Number1
Product Code DSP
UDI-Device Identifier30801902084966
UDI-Public30801902084966
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K162820
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 02/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/11/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberIPN001112
Device Catalogue NumberIAP-0700
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/07/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-