Model Number 1012582-16
Medical Device Problem Codes Improper or Incorrect Procedure or Method (2017); Patient-Device Incompatibility (2682)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/16/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no similar incidents from this lot.It was reported that the graftmaster was deployed with a maximum pressure of 20 atmospheres (atm).It should be noted the graftmaster rapid exchange (rx), coronary stent graft system, domestic, instructions for use (ifu) specifies the rated burst pressure (rbp) is 16 atm and clearly states not to exceed the rbp.The investigation was unable to determine a conclusive cause for the reported failure to seal.There is no indication of a product quality issue with respect to manufacture, design or labeling.
 
Event or Problem Description
It was reported this was a procedure to treat a perforated location in the right coronary artery.It was noted this was a st-elevation myocardial infarction (stemi) patient with severe three vessel disease.A perforation to right coronary artery occurred while ballooning vessel.A 4.00x16 graftmaster was inserted, inflated to 20 atmospheres (atms) and implanted.However, the perforation was unable to be sealed.The patient then passed away, but the physician stated the implanted graftmaster did not cause or contribute to the patient death.There as no clinically significant delay in the procedure.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GRAFTMASTER RX CORONARY STENT GRAFT SYSTEM
Common Device Name
CORONARY STENT DELIVERY SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005718570 (P099)
cashel road
clonmel tipperary
EI  
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key13740259
Report Number2024168-2022-02547
Device Sequence Number4762952
Product Code MAF
UDI-Device Identifier08717648176449
UDI-Public08717648176449
Combination Product (Y/N)N
Initial Reporter StateAR
Initial Reporter CountryUS
PMA/510(K) Number
H000001
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2023
Device Model Number1012582-16
Device Catalogue Number1012582-16
Device Lot Number1052541
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/17/2022
Initial Report FDA Received Date03/11/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/25/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexFemale
-
-