• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: W. L. GORE & ASSOCIATES, INC. GORE® EXCLUDER® AAA ENDOPROSTHESIS; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

W. L. GORE & ASSOCIATES, INC. GORE® EXCLUDER® AAA ENDOPROSTHESIS; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT Back to Search Results
Model Number PLC141200
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code Aneurysm (1708)
Date of Event 02/11/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Patient medical history includes but is not limited to: aaa, pvd, hypertension.The instructions for use (ifu) for the gore® excluder® aaa endoprosthesis states, adverse events that may occur and / or require intervention include but are not limited to: endoprosthesis: migration, dilatation, endoleak.
 
Event or Problem Description
On (b)(6) 2018, this patient underwent endovascular treatment for an abdominal aortic aneurysm and was implanted with gore® gore® excluder® aaa endoprostheses featuring c3® delivery system.The patient tolerated the procedure.On (b)(6), 2022, the patient underwent follow-up computed tomography angiography wherein the physician identified a type 1b endoleak was present and decided to treat it.The physician decided to extend the original device in the right common iliac artery (rcia) as it had migrated approximate 4cm proximally due to disease progression (dilation of the rcia) distal to the original device.Two additional contralateral leg components were implanted to seal the rcia.In addition, the physician also extended the original graft on the left side with a contralateral leg component to ensure long-term seal although no type 1b endoleak was identified on the left side.The endoleak resolved, and the patient tolerated the procedure well.Patient is reported to be doing fine.
 
Additional Manufacturer Narrative
H6: code 213- a review of the manufacturing records for the device(s) verified that the lot(s) met all pre-release specifications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GORE® EXCLUDER® AAA ENDOPROSTHESIS
Common Device Name
SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
Manufacturer (Section D)
W. L. GORE & ASSOCIATES, INC.
1505 n. fourth street
flagstaff AZ 86004
Manufacturer (Section G)
MEDICAL SILICON VALLEY B/P
2890 de la cruz blvd.
santa clara CA 95050
Manufacturer Contact
laura crawford
1505 n. fourth street
flagstaff, AZ 86004
9285263030
MDR Report Key13742228
Report Number3013164176-2022-01328
Device Sequence Number4608048
Product Code MIH
UDI-Device Identifier00733132618675
UDI-Public00733132618675
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
P020004
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/25/2021
Device Model NumberPLC141200
Device Catalogue NumberPLC141200
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/11/2022
Supplement Date Received by Manufacturer02/11/2022
Initial Report FDA Received Date03/12/2022
Supplement Report FDA Received Date03/16/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/26/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Required Intervention; Hospitalization;
Patient Age76 YR
Patient SexMale
Patient Weight79 KG
-
-