• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARROW INTERNATIONAL LLC ARROW AGBA PICC/DELTA KIT: 1-L 4.5 FR X 55 CM; CATHETER INTRAVASCULAR THERAPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARROW INTERNATIONAL LLC ARROW AGBA PICC/DELTA KIT: 1-L 4.5 FR X 55 CM; CATHETER INTRAVASCULAR THERAPE Back to Search Results
Model Number IPN037306
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code Device Embedded In Tissue or Plaque (3165)
Date of Event 01/28/2022
Type of Reportable Event Serious Injury
Event or Problem Description
The complaint is reported as: part of the peelaway sheath broke off in the patient's subcutaneous tissue upon removal.The piece of sheath has remained in the patient's arm since the insertion on 28-january.Removal of the sheath piece was to take place yesterday or today 18-february by cut down of the subcutaneous tissues at the insertion site.The piece was not actually in the vessel.The patient's condition is reported as fine.
 
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
The complaint is reported as: part of the peel away sheath broke off in the patient's subcutaneous tissue upon removal.The piece of sheath has remained in the patient's arm since the insertion on (b)(6).Removal of the sheath piece was to take place yesterday or today on (b)(6) by cut down of the subcutaneous tissues at the insertion site.The piece was not actually in the vessel.The patient's condition is reported as fine.
 
Additional Manufacturer Narrative
(b)(4).Complaint verification testing could not be performed as it was reported that the sample is not available for return.A device history record review was performed, and a relevant finding was identified.A non-conformance was initiated for batch: 34c21j0107 to address the issue of a damaged sheath body.Despite this, it cannot be confirmed if the two issues are linked without the sample or an image to confirm the reported damage.Without the device to evaluate the complaint could not be confirmed and the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARROW AGBA PICC/DELTA KIT: 1-L 4.5 FR X 55 CM
Common Device Name
CATHETER INTRAVASCULAR THERAPE
Manufacturer (Section D)
ARROW INTERNATIONAL LLC
morrisville NC
Manufacturer (Section G)
ARROW INTERNATIONAL DE MEXICO S.A. DE C.V.
ave. washington 3701
colonia panamericana, chihuahua
chihuahua 31200
MX   31200
Manufacturer Contact
katharine tarpley
3015 carrington mill blvd
morrisville, NC 27560
MDR Report Key13742509
Report Number9680794-2022-00152
Device Sequence Number4544738
Product Code LJS
UDI-Device Identifier50801902120781
UDI-Public50801902120781
Combination Product (Y/N)N
Initial Reporter StateKS
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Expiration Date02/28/2023
Device Model NumberIPN037306
Device Catalogue NumberCDC-45541-VPS2
Device Lot Number13F21K0618
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/17/2022
Supplement Date Received by Manufacturer04/04/2022
Initial Report FDA Received Date03/12/2022
Supplement Report FDA Received Date04/04/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/28/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
N/A.; N/A.
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
-
-