| Model Number |
1352282 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/17/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
This is 2 of 3 reports linked to mfg report numbers: 3014334038-2022-00049, 3014334038-2022-00050.A facility reported that the blade o/r table bracket (1352282) broke in the middle of a tympanoplasty procedure.All parts were recovered and were taken off the surgical field; no fragments fell into the surgical site.It was reported that an x-ray was performed to assure that no parts were left in the patient following the event.The event did not lead to patient injury, death, or delay in procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
N/a.
|
| |
|
Additional Manufacturer Narrative
|
|
The blade o/r table bracket (1352282) was returned for evaluation: failure analysis - evaluation was unable to conclusively verify customer information as valid.The bracket was undamaged and no defect was found.Testing was within specification.Root cause - the returned 1352282 table bracket is in used condition with no visible defect.The bracket was in use with a broken 1352275 speculum arm.The arm is being investigated under a separate complaint.No further investigation required based on the acceptability of risk and no adverse trends identified.This will be monitored and trended going forward.
|
| |
|
Search Alerts/Recalls
|