• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER TRAUMA KIEL TROCHANTERIC NAIL KIT, TI GAMMA3® Ø10X170MM X 125°; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER TRAUMA KIEL TROCHANTERIC NAIL KIT, TI GAMMA3® Ø10X170MM X 125°; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Model Number 3125-0170S
Medical Device Problem Code Mechanical Jam (2983)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/15/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Upon completion of the investigation, additional information will be provided in a supplemental report.
 
Event or Problem Description
As reported; "the target device and the holding screw were stuck and could not be removed from the nail after the nail insertion.After that, other instruments and nail were used and the surgery was finished." time extension of about 20 minutes.
 
Additional Manufacturer Narrative
The reported event could be confirmed, since the device was returned, and matches the alleged failure.The device inspection revealed the following: the received device came back jammed/assembled which was unblocked by holding in wise after that there were no visible damages, but a huge amount of dry blood was observed which could be a reason for the jam issue.After the decontamination, the functional test was performed by assembling and disassembling the device in which there was no issue observed just the chunks of dried blood/substance fell off.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.Based on the investigation, the root cause was attributed to a user related issue.If the device is returned or if any additional information is provided, the investigation will be reassessed.
 
Event or Problem Description
As reported; "the target device and the holding screw were stuck and could not be removed from the nail after the nail insertion.After that, other instruments and nail were used and the surgery was finished." time extension of about 20 minutes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TROCHANTERIC NAIL KIT, TI GAMMA3® Ø10X170MM X 125°
Common Device Name
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
GM  D-24232
Manufacturer (Section G)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
GM   D-24232
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key13743544
Report Number0009610622-2022-00090
Device Sequence Number17259679
Product Code HSB
UDI-Device Identifier04546540375063
UDI-Public04546540375063
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K034002
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2026
Device Model Number3125-0170S
Device Catalogue Number31250170S
Device Lot NumberK0C46B4
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/10/2022
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 02/15/2022
Supplement Date Received by Manufacturer06/08/2022
Initial Report FDA Received Date03/12/2022
Supplement Report FDA Received Date07/06/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/22/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-