ST PAUL BCI; BLOOD PRESSURE MONITORING Back to Search Results
Model Number 6004001
Medical Device Problem Code Power Problem (3010)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a smiths medical patient monitoring|bci blood pressure monitor mini-torr plus - 6004 unit was not turning on fully.No patient injury was reported.
 
Additional Manufacturer Narrative
Event problem and evaluation codes: updated after device evaluation.Device evaluation: the device was returned for investigation.A visual inspection and functional test were performed.Visual inspection found the on off switch was used to the point the paint came off.The customer stated problem was not duplicated.No fault found with the device regarding the reported customer problem.The cause of the reported problem could not be determined.The dhr review is not applicable or relevant because the results of the complaint investigation do not indicate a problem with the initial manufacture or prior repair of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BCI
Common Device Name
BLOOD PRESSURE MONITORING
Manufacturer (Section D)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer (Section G)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer Contact
david halverson
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key13743716
Report Number3012307300-2022-04478
Device Sequence Number4838906
Product Code DQA
UDI-Device Identifier10610586036583
UDI-Public10610586036583
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
K031742
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 03/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number6004001
Device Catalogue Number6004001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/29/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/26/2021
Initial Report FDA Received Date03/12/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/20/2004
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1