MAUDE Adverse Event Report: PENUMBRA, INC. PENUMBRA SYSTEM 3MAX REPERFUSION CATHETER; NRY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PENUMBRA, INC. PENUMBRA SYSTEM 3MAX REPERFUSION CATHETER; NRY Back to Search Results
Model Number 3MAXC-B
Medical Device Problem Codes Break (1069); Physical Resistance/Sticking (4012)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/14/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This device is available for return.A follow up mdr will be submitted upon completion of the device investigation.
 
Event or Problem Description
The patient was undergoing a thrombectomy procedure in the m1 segment of the middle cerebral artery (mca) using a penumbra system 3max reperfusion catheter (3maxc), benchmark bmx96 access system (benchmark max), and a penumbra system red 72 reperfusion catheter (red72).During the procedure, the physician completed one pass using the benchmark max, red72 and 3maxc.While retracting the 3maxc, the physician experienced resistance and upon removal, the physician found the 3maxc broken at the distal end and hanging by a thread.Therefore, the 3maxc was not used for the remainder of the procedure.The procedure was completed using a new 3maxc, the same benchmark max and the same red72.There was no report of an adverse effect to the patient.
 
Additional Manufacturer Narrative
Please note that the device associated with this complaint was expected to be returned; however, additional information received from the penumbra sales representative indicated that the device was discarded and is no longer available for return.Therefore, without the return of the device, the root cause of the problem cannot be determined.Based on this information, corrections were made to: 1.Section d.Box 10.Device available for evaluation? (do not send to fda) 2.Section h.Box 3.Reason for non-evaluation 3.Section h.Box 3.''other'' reason for non-evaluation the manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.H3 other text : placeholder.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENUMBRA SYSTEM 3MAX REPERFUSION CATHETER
Common Device Name
NRY
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
Manufacturer Contact
veronica farris
one penumbra place
alameda, CA 94502
5107483200
MDR Report Key13744879
Report Number3005168196-2022-00118
Device Sequence Number4544212
Product Code NRY
UDI-Device Identifier00814548019956
UDI-Public00814548019956
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
K160449
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/01/2022
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/14/2024
Device Model Number3MAXC-B
Device Catalogue Number3MAXC
Device Lot NumberF110030
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/14/2022
Supplement Date Received by Manufacturer03/17/2022
Initial Report FDA Received Date03/12/2022
Supplement Report FDA Received Date04/01/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/15/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
Date Report Sent to FDA01/01/2005
-
-