• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZOLL CIRCULATION AUTOPULSE® RESUSCITATION MODEL 100; CARDIAC CHEST COMPRESSOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZOLL CIRCULATION AUTOPULSE® RESUSCITATION MODEL 100; CARDIAC CHEST COMPRESSOR Back to Search Results
Model Number MODEL 100
Medical Device Problem Code Use of Device Problem (1670)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/23/2021
Type of Reportable Event Malfunction
Event or Problem Description
The autopulse platform (sn unknown) was used to resuscitate a patient in cardiac arrest.The user installed the new lifeband (lot# unknown) to use with the platform, however, the platform could not make compressions.The crew reverted to manual cpr for the rest of the call.No consequences or impact to the patient.Upon returning to the ems center, the crew noticed that the lifeband band clip that snaps into the autopulse platform was broken.Please see the following related mfr report: mfr #3010617000-2022-00249 for the autopulse lifeband.
 
Additional Manufacturer Narrative
The device associated with this complaint will not be returned for evaluation.After the call, the autopulse platform was checked and the lifeband was observed with the broken band clip.The platform was verified to be functional and have been used by the user without issues.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOPULSE® RESUSCITATION MODEL 100
Common Device Name
CARDIAC CHEST COMPRESSOR
Manufacturer (Section D)
ZOLL CIRCULATION
2000 ringwood ave.
san jose CA 95131
Manufacturer (Section G)
ZOLL CIRCULATION
2000 ringwood ave.
san jose CA 95131
Manufacturer Contact
kim thoa nguyen
2000 ringwood ave,
san jose, CA 95131
4084192922
MDR Report Key13749163
Report Number3010617000-2022-00317
Device Sequence Number20076239
Product Code DRM
UDI-Device Identifier00849111000550
UDI-Public00849111000550
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K112998
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 03/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberMODEL 100
Device Catalogue Number8700-0730-02
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/17/2022
Initial Report FDA Received Date03/12/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-