ALCON LABORATORIES IRELAND LTD. ACRYSOF IQ ASPHERIC IOL WITH THE ULTRASERT PRE-LOADED DELIVERY SYSTEM; INTRAOCULAR LENS Back to Search Results
Model Number AU00T0
Medical Device Problem Codes Device Damaged by Another Device (2915); Failure to Eject (4010)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/26/2022
Type of Reportable Event Malfunction
Event or Problem Description
A nurse reported during the intraocular lens implantation on the third try the lens advanced smoothly and the lens was implanted.The nurse thought she had seen small scratch on the surface of the lens.It was also instructed that the lens should not be used if the plunger goes above the lens.There was contact with the patient and the lens remained implanted.There was no patient harm and hospitalization or intervention.The surgery was completed on the same day with the same lens, no replacement lens was used.
 
Additional Manufacturer Narrative
A sample device was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.There have been no other complaints reported in the lot number.Root cause has not been identified.The manufacturer internal reference number is: (b)(4).
 
Additional Manufacturer Narrative
The complainant has indicated temperature discrepancy: while we are unable to determine the origin of the reported complaint, our observations suggest that based on the information provided by the customer, the most likely root cause is failure to follow the ifu, as the customer states that temperature at the time of the procedure more than 23°c/ 73°f.If the operating room temperature is too high (> 23°c / 73° f) lens folding consistency is negatively affected, the lens is more adherent and this may inhibit lens advancement.In addition to this, the file also states that the customer has been instructed to abort the device if plunger override occurs.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF IQ ASPHERIC IOL WITH THE ULTRASERT PRE-LOADED DELIVERY SYSTEM
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
ALCON LABORATORIES IRELAND LTD.
cork business&technology park
model farm road
cork 00000
EI  00000
Manufacturer (Section G)
ALCON LABORATORIES IRELAND LTD.
cork business&technology park
model farm road
cork 00000
EI   00000
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key13751376
Report Number9612169-2022-00110
Device Sequence Number14604317
Product Code HQL
UDI-Device Identifier00380652394888
UDI-Public00380652394888
Combination Product (Y/N)N
Initial Reporter CountryFI
PMA/510(K) Number
P930014
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 06/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2023
Device Model NumberAU00T0
Device Lot Number21282988
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/17/2022
Supplement Date Received by Manufacturer05/10/2022
Initial Report FDA Received Date03/13/2022
Supplement Report FDA Received Date06/02/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/09/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DISCOVISC OPHTHALMIC VISCOSURGICAL DEVICE; MONARCH IOL DELIVERY SYSTEM, CARTRIDGE; UNSPECIFIED DISCOVISC; UNSPECIFIED MONARCH CARTRIDGES
Patient SexFemale
-
-