ARTHROCARE CORP. AMBIENT HIPVAC 50 IFS; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number 72290004
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/21/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Internal complaint reference case (b)(4).
 
Event or Problem Description
It was reported that during an arthroscopy, a part of the black plastic insulation of the ambient broke away.The pieces were removed by grasper.The procedure was completed with a s+n back-up device.Non-significant delay was reported, and no other complications were reported.
 
Additional Manufacturer Narrative
H3, h6: the reported device was received for evaluation.A visual inspection of the returned instrument shows no manufacturing abnormalities.The insulation has a small piece peeled away.Product was out of the original packaging.No packaging returned.A functional evaluation revealed the controller generated an e7 error when the wand was connected.When used with a bypass box the device generated plasma and coagulation as intended.Suction performed.It was determined the device contributed to the reported event.The complaint was confirmed and the root cause was associated with unintended use of the device.Factors that could have contributed to the reported event include: 1) mechanical displacement of tissue through applied force or using the device as a lever to enlarge a surgical site or gain access to tissue.2) damage to the tip and/or shaft insulation that occurred as a result of contact with metal objects while activating the wand.3) contacting the distal portion of cannula with the shaft when inserting and removing the wand.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A complaint history review found no similar reported events.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.No containment or corrective actions are recommended at this time.Our clinical investigation concluded: a clinical review states this case reports, during an arthroscopy, ¿a part of the black plastic insulation broke away.¿ the pieces were reportedly removed by graspers and the procedure was completed with a smith and nephew back-up device and a non-significant delay.No other complications were reported.Per case communication, ¿it¿s not 100% sure if all pieces were removed.¿ no additional clinically relevant information was provided.Based on the information provided, the root cause of the reported ¿part of the black plastic insulation broke awake¿ cannot be definitively concluded.The broken instrument was reportedly removed with a grasper, however, ¿it¿s not 100% sure if all pieces were removed.¿ the patient impact beyond the reported cannot be determined.Therefore, no further clinical medical assessment can be rendered.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMBIENT HIPVAC 50 IFS
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
ARTHROCARE CORP.
7000 w. william cannon
austin TX 78735
Manufacturer (Section G)
ARTHROCARE CORP.
7000 w. william cannon
austin TX 78735
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key13756824
Report Number3006524618-2022-00092
Device Sequence Number17635772
Product Code GEI
UDI-Device Identifier00885556614686
UDI-Public00885556614686
Combination Product (Y/N)N
Initial Reporter CountrySZ
PMA/510(K) Number
K161481
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/05/2024
Device Model Number72290004
Device Catalogue Number72290004
Device Lot Number2073768
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/22/2022
Supplement Date Received by Manufacturer06/02/2022
Initial Report FDA Received Date03/14/2022
Supplement Report FDA Received Date06/06/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/05/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-