Model Number N/A
Device Problems Positioning Failure (1158); Fracture (1260)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/03/2022
Event Type  malfunction  
Event Description
It was reported that the surgeon screwed the shell onto the cup inserter, hammered it in, saw that it was loose, pulled it out and discovered a piece of thread was torn from the inserter and remained in the shell.No delay and surgical technique was utilized.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Manufacturer Narrative
(b)(4).Report source: foreign: (b)(6).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 0001822565-2022-00709.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Straight shell inserter item 00780401520 lot 63037846 was returned for review against the complaint.Visual review of the shell inserter shows wear and tear from previous uses.Nicks, gouges, and scratches to the handle and body.The strike plate shows indentations.The threaded tip is confirmed to be fractured.Device has an approximate field age of 7 years.Device history record was reviewed and no discrepancies relevant to the reported event were found.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STRAIGHT SHELL INSERTER
Type of Device
INSTRUMENT, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key13756887
MDR Text Key289207152
Report Number0001822565-2022-00708
Device Sequence Number1
Product Code LXH
Combination Product (y/n)N
Reporter Country CodePL
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 07/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/14/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number00780401520
Device Lot Number63037846
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received07/13/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/30/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Removal/Correction NumberN/A
Patient Sequence Number1
Treatment
CAT#: 00875705002, 50MM SHELL, LOT#: 64932306.
-
-