• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. BIO-COMPS-ITE CRKSCRW 5.5X14.7MM; FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. BIO-COMPS-ITE CRKSCRW 5.5X14.7MM; FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE Back to Search Results
Model Number BIO-COMPS-ITE CRKSCRW 5.5X14.7MM
Date of Event 02/22/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
 
Event or Problem Description
On (b)(6) 2022 it was reported by a sales representative via sems that after normal insertion of tripleplay anchor and passing 3 limbs of suture there were no obvious problems with the ar-1927bcf-3 anchor.The surgeon then proceeded to tension the limbs in order to tie down the repair.While the surgeon began to pull both limbs of the blue suture the sales rep, noticed how it was pulling the tear up significantly.Both the surgeon and the sales rep noticed how the blue sutures weren¿t in the anchor anymore and had come free from the eyelet.The surgeon did not unload the anchor.It popped through the material eyelet.The sales rep asked him to lightly pull the solid whites and the same thing happened and once more with the stripes.The anchor remained fully seated in the bone but the eyelet was not holding any of the sutures not allowing us to use it for repair.This was discovered during use in a rotator cuff repair procedure.Additional information received 02/23/2022.On (b)(6) 2022 it was reported by a sales representative via email that after normal insertion of tripleplay anchor and passing 3 limbs of suture using the ar-1927bcf-3 anchor.The surgeon proceeded to tension the limbs to tie down the repair.While the surgeon began to pull both limbs of the blue suture the sales rep, noticed how it was pulling the tear up significantly.Both the surgeon and the sales rep noticed the blue sutures weren¿t in the anchor anymore and had come free from the eyelet.The surgeon did not unload the anchor.The surgeon lightly pulls the solid whites and the same thing happened and once more with the stripes.The anchor remained fully seated in the bone, but the eyelet was not holding any of the sutures not allowing us to use it for repair.This was discovered during use in a rotator cuff repair procedure. the case was completed by using the silver double lined punch ar-1927pb.The surgeon punched to the first laser line (ft) before anchor insertion.After the corkscrew sutures pulled out, we then used a fibertape and 4.75 swivelock to perform a speed fix.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIO-COMPS-ITE CRKSCRW 5.5X14.7MM
Common Device Name
FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
vik bajnath
8009337001
MDR Report Key13757294
Report Number1220246-2022-04588
Device Sequence Number11361570
Product Code MAI
UDI-Device Identifier00888867023369
UDI-Public00888867023369
Combination Product (Y/N)N
Initial Reporter StateLA
Initial Reporter CountryUS
PMA/510(K) Number
K082810
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial
Report Date (Section B) 03/14/2022
1 Device was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberBIO-COMPS-ITE CRKSCRW 5.5X14.7MM
Device Catalogue NumberAR-1927BCF-3
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/22/2022
Initial Report FDA Received Date03/14/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
-
-