STRYKER ORTHOPAEDICS-MAHWAH TRIDENT 0 DEG INSERT 36MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED Back to Search Results
Model Number 623-00-36D
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 02/16/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.Review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other similar events for the lot referenced.The following devices were also listed in this report: tridentii tritanium cluster50d; cat#702-04-50d ; lot#84664101a, size 3 accolade ii 127 deg; cat#6721-0330 ; lot#87370402, delta v-40 ceramic head 36/0; cat#6570-0-136 ; lot#88992802,and 6.5mm low profile hex screw 30mm; cat#7030-6530 ; lot#ynna2.It cannot be determined which, if any of these devices may have caused or contributed to the patient's experience.
 
Event or Problem Description
The patient is under infection after total hip arthroplasty.The operation date was (b)(6)2021 and he visited the hospital again with infection two months later.Left leg.Product: stem(accolade ii),head(v40 ceramic),insert(x3 poly insert),cup(trident ii tritanium cluster),hip screw(6.5 low profile hex).
 
Additional Manufacturer Narrative
An event regarding infection involving a trident liner was reported.The event was not confirmed.Method & results: -product evaluation and results: material analysis, visual, functional and dimensional inspections could not be performed as the device remains implanted.-clinician review: insufficient information was provided to support a medical review.-product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.-complaint history review: there have been no other similar events for the lot or sterile lot referenced.Conclusions: all stryker products sold as sterile are validated to a minimum sterility assurance level sal of 10^-6 in accordance with applicable iso standards.The exact cause of the event could not be determined because insufficient information was provided.Further information such as return of the device, pathology reports, pre- and post-operative x-rays and the primary operative report as well as patient history and follow-up notes are needed to complete the investigation for determining root cause.No further investigation for this event is possible at this time.If devices and/or additional information become available to indicate further evaluation is warranted, this record will be reopened.
 
Event or Problem Description
The patient is under infection after total hip arthroplasty.The operation date was (b)(6) 2021 and he visited the hospital again with infection two months later.Left leg.-product: stem (accolade ii), head (v40 ceramic), insert (x3 poly insert), cup trident ii tritanium cluster), hip screw (6.5 low profile hex).Update: aspiration was performed and iv antibiotics were administered 3 weeks post operatively due to elevated crp.No bacterial pathogen was confirmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIDENT 0 DEG INSERT 36MM
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer Contact
brad curtis
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key13757912
Report Number0002249697-2022-00383
Device Sequence Number14938142
Product Code LPH
UDI-Device Identifier07613327025309
UDI-Public07613327025309
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
K062419
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/31/2026
Device Model Number623-00-36D
Device Catalogue Number623-00-36D
Device Lot Number9E4380
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/17/2022
Supplement Date Received by Manufacturer08/02/2022
Initial Report FDA Received Date03/14/2022
Supplement Report FDA Received Date08/26/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/09/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Other; Required Intervention;
Patient Age82 YR
Patient SexFemale
-
-