ZIMMER BIOMET, INC. COMP AUG MINI BSPLT W TPR LG; PROSTHESIS, SHOULDER Back to Search Results
Catalog Number 110032430
Medical Device Problem Code Migration or Expulsion of Device (1395)
Health Effect - Clinical Code Failure of Implant (1924)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Common device name: phx.Concomitant medical products: comp rvs cntrl 6.5x25mm st/rst: cat# 115395, lot# 071970; comp lk scr 3.5hex 4.75x30 st: cat# 180553, lot# 420090; comp lk scr 3.5hex 4.75x30 st: cat# 180553, lot# 498970; comp lk scr 3.5hex 4.75x15 st: cat# 180550, lot# 234020; unknown locking screw: cat#ni, lot#ni.The device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 0001825034-2022-00539, 0001825034-2022-00540, 0001825034-2022-00541, 0001825034-2022-00542, 0001825034-2022-00543.
 
Event or Problem Description
It was reported that a patient is being considering for a revision at an unknown date due to total failure of the glenoid components (broken screws and baseplate pulled out from glenoid).Attempts have been made and additional information on the reported event is unavailable at this time.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.Review of the device history record(s) identified no deviations or anomalies during manufacturing.Radiographs were provided and reviewed by a health care professional.Review of the available records identified the following: fractured glenosphere screws with loosening of the glenoid component.Possibly related to extensive glenoid heterotopic ossification.Radiolucency surrounding the glenosphere, consistent with loosening.Bony impaction of the humeral component with the heterotopic ossification arising of the inferior aspect of the glenoid.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information is available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COMP AUG MINI BSPLT W TPR LG
Common Device Name
PROSTHESIS, SHOULDER
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key13758034
Report Number0001825034-2022-00538
Device Sequence Number14909484
Product Code PHX
UDI-Device Identifier00880304993259
UDI-Public(01)00880304993259(17)231102(10)64228644
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K172502
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/02/2023
Device Catalogue Number110032430
Device Lot Number64228644
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/24/2022
Supplement Date Received by Manufacturer04/21/2022
Initial Report FDA Received Date03/14/2022
Supplement Report FDA Received Date04/26/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/02/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexMale
-
-