• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AMO PUERTO RICO MFG. INC. TECNIS IOL; INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AMO PUERTO RICO MFG. INC. TECNIS IOL; INTRAOCULAR LENS Back to Search Results
Model Number PCB00
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/11/2022
Type of Reportable Event Malfunction
Event or Problem Description
It has been reported defective (the plunger has broken the haptic).No contact with patient's eye.No further information has been provided.
 
Additional Manufacturer Narrative
Per eu regulation 2016/679 (general data protection regulation), patient identifiers were not collected or recorded and therefore are not available.Implant date: if implanted, give date: lens was not implanted explant date: if explanted, give date: lens was not implanted; therefore not explanted reporter telephone no.: (b)(6).Device evaluated by mfr: the device is not received; therefore, a failure analysis of the complaint device cannot be completed.A review of the device/lot history record and complaint trending for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Additional Manufacturer Narrative
Additional information: section d9.Device available for evaluation? yes returned to manufacturer on: 1 march 2021 section h3.Device evaluated by manufacturer? yes device evaluation: visual inspection under magnification revealed that the complaint lens was received with a detached haptic.The lens was cleaned and, no further issues were observed with the lens.Visual inspection of the handpiece revealed that the handpiece was received with the plunger rod in the advanced and locked position.Further observation revealed that the plunger rod tip was damaged.The external assembly was inspected and, no further issues were identified.Based on the definition of "haptic damaged" the complaint issue could not be confirmed.However, the observed condition of "haptic detached" is similar to the complaint issue "haptic damaged" but based on the complaint investigation results the observed issue could not be confirmed to be related to the manufacturing or design process.Therefore, there is no indication of a product deficiency or product malfunction.Conclusion: as a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
AMO PUERTO RICO MFG. INC.
road 402 north, anasco ind. pk
anasco PR 00610
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
7142478552
MDR Report Key13758267
Report Number3012236936-2022-00626
Device Sequence Number2455868
Product Code HQL
UDI-Device Identifier05050474558328
UDI-Public(01)05050474558328(17)241007
Combination Product (Y/N)N
Initial Reporter CountrySP
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 05/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/07/2024
Device Model NumberPCB00
Device Catalogue NumberPCB0000230
Was Device Available for Evaluation? Device Returned to Manufacturer
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/15/2022
Supplement Date Received by Manufacturer04/20/2022
Initial Report FDA Received Date03/14/2022
Supplement Report FDA Received Date05/10/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/07/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-