CORDIS CORPORATION SI AVANTI+ 8F STD W/GW NO OBT; DILATOR, VESSEL, FOR PERCUTANEOUS CATHETERIZATION Back to Search Results
Model Number 504608X
Medical Device Problem Codes Partial Blockage (1065); Detachment of Device or Device Component (2907)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 11/17/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Additional information is pending and will be submitted within 30 days upon receipt.
 
Event or Problem Description
As reported, the avanti plus catheter sheath introducer (si avanti+ 8f std w/gw no obt) fell off (separated) from a patient and there was blood return.The patient's heart rate and blood pressure did not change at that time and small clips were used to clip the arterial collaterals and reflex the catheter to prevent bleeding.The patient was reported to have been initially admitted due to acute coronary syndrome and was placed in an intra-aortic balloon pump (iabp).The pump was subsequently replaced with a fixed catheter (the avanti plus catheter) which was found disconnected from the patient at bedtime.The collateral catheter was immediately reversed, and the event was reported to the doctor and head nurse.According to the additional information provided, the patient was reported to have been hospitalized due to acute myocardial infarction and acute thromboembolism and underwent a percutaneous coronary intervention (pci) via the anterior descending branch.The lesion was not calcified; there was no vessel tortuosity and ninety-percent stenosis.There were no anomalies noted when the product was removed from the package.The product was stored and handled according to the instructions for use (ifu).The product was inspected and prepped according to the ifu and there was no difficulty experienced when prepping the device.The patient did not require extended hospitalization as a result of this event.There was resistance during withdrawal of the sheath from the vessel.The physician removed the sheath and checked the inner diameter of the sheath to confirm if it was the same throughout the entire sheath, but there was a bulge (obstruction) which was believed to be the cause of the resistance.The reported separation occurred during procedural use.The product was discarded at site.
 
Additional Manufacturer Narrative
As reported, the avanti plus catheter sheath introducer (si avanti+ 8f std w/gw no obt) fell off (separated) from a patient and there was blood return.The patient's heart rate and blood pressure did not change at that time and small clips were used to clip the arterial collaterals and reflex the catheter to prevent bleeding.The patient was reported to have been initially admitted due to acute coronary syndrome and was placed in an intra-aortic balloon pump (iabp).The pump was subsequently replaced with a fixed catheter (the avanti plus catheter) which was found disconnected from the patient at bedtime.The collateral catheter was immediately reversed, and the event was reported to the doctor and head nurse.According to the additional information provided, the patient was reported to have been hospitalized due to acute myocardial infarction and acute thromboembolism and underwent a percutaneous coronary intervention (pci) via the anterior descending branch.The lesion was not calcified; there was no vessel tortuosity and ninety-percent stenosis.There were no anomalies noted when the product was removed from the package.The product was stored and handled according to the instructions for use (ifu).The product was inspected and prepped according to the ifu and there was no difficulty experienced when prepping the device.The patient did not require extended hospitalization because of this event.There was resistance during withdrawal of the sheath from the vessel.The physician removed the sheath and checked the inner diameter of the sheath to confirm if it was the same throughout the entire sheath, but there was a bulge (obstruction) which was believed to be the cause of the resistance.The reported separation occurred during procedural use.The product was not returned for analysis as it was discarded.A product history record (phr) review of lot 18035168 revealed no anomalies or non-conformances during the manufacturing and inspection processes that can be associated with the reported event.Without the return of the device or images for analysis, the reported customer event ¿catheter sheath introducer- separated - in patient¿ and ¿catheter sheath introducer - obstructed - in patient¿ could not be confirmed.Procedural/handling factors may have contributed to the reported event.Users are trained to inspect for signs of damage prior to and during use.Any product with damage is not to be used.Information for safety is provided in the products labeling with the intent to make the user aware of the risks.According to the instructions for use (ifu), although not intended as a mitigation of risk, ¿aspirate from the sideport extension to remove any potential air.After aspiration and flushing, consider establishing a heparinized solution or suitable isotonic solution via the sideport extension.The addition of a heparinized saline drip via the sideport can help in the prevention of thrombus formation.Hold the sheath in place when inserting, positioning, or removing the catheter.The suture collar may be used as a temporary suture site.¿ based on the information available and the phr review, there is no indication that the event is related to the device design or manufacturing process.Therefore, no preventative or corrective actions will be taken at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SI AVANTI+ 8F STD W/GW NO OBT
Common Device Name
DILATOR, VESSEL, FOR PERCUTANEOUS CATHETERIZATION
Manufacturer (Section D)
CORDIS CORPORATION
14201 nw 60th avenue
miami lakes FL 33014 2802
Manufacturer (Section G)
CORDIS CORPORATION
14201 nw 60th avenue
miami lakes FL 33014 2802
Manufacturer Contact
karla castro
14201 nw 60th avenue
miami lakes, FL 33014-2802
7863138372
MDR Report Key13758832
Report Number9616099-2022-05448
Device Sequence Number2455907
Product Code DRE
UDI-Device Identifier10705032010153
UDI-Public10705032010153
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K970392
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date06/30/2024
Device Model Number504608X
Device Catalogue Number504608X
Device Lot Number18035168
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 02/15/2022
Supplement Date Received by Manufacturer03/12/2022
Initial Report FDA Received Date03/14/2022
Supplement Report FDA Received Date03/22/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/03/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNK CATHETER
Outcome Attributed to Adverse Event Life Threatening;
Patient SexUnknown
-
-