| Model Number |
70914-70 |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Codes |
Hyperglycemia (1905); Loss of consciousness (2418)
|
| Date of Event |
02/14/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
At this time product has not yet been returned.An extended investigation has been performed for the reported complaint and it has been determined that there was no indication that the product did not meet specification.The dhrs (device history record) for the freestyle freedom lite meter were reviewed and the dhrs showed the freestyle freedom lite meter passed all tests prior to release.The dhrs (device history record) for the freestyle strips were reviewed and the dhrs showed the freestyle strips passed all tests prior to release.Retain testing was performed for freestyle strips and all units performed in specification and passed.If the product is returned, a physical investigation will be performed and a follow-up report submitted.The date the incident occurred is unknown.The date entered in is the date abbott diabetes care became aware of the event.All pertinent information available to abbott diabetes care has been submitted.
|
| |
|
Event or Problem Description
|
|
Customer reported receiving unspecified readings that were erratic on adc device and he experienced a loss of consciousness.Customer was seen by a healthcare provider and a reading of 380 mg/dl was obtained.Customer was diagnosed with hyperglycemia and received unspecified treatment.There was no report of death or permanent impairment associated with this event.
|
| |
|