• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SISTEMAS MEDICOS ALARIS, S.A. DE C.V. BD ALARIS¿ SMARTSITE¿ GRAVITY BURETTE SET; INTRAVASCULAR ADMINISTRATION SET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SISTEMAS MEDICOS ALARIS, S.A. DE C.V. BD ALARIS¿ SMARTSITE¿ GRAVITY BURETTE SET; INTRAVASCULAR ADMINISTRATION SET Back to Search Results
Model Number 10037032
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/23/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that 5 bd alaris¿ smartsite¿ gravity burette sets leaked fluid during use.The following information was provided by the initial reporter: "customer has confirmed that the burette is leaking fluid during the use.".
 
Additional Manufacturer Narrative
A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Additional Manufacturer Narrative
H6: investigation summary a 10037032 product was not available for investigation, however the customer confirmed that the complaint sample was from lot 21109297.As part of the feedback the customer provided a video of the affected product, analysis of the video confirmed the customer's experience with leakage observed from the base of the burette.The details of this feedback were forwarded to the manufacturing site for investigation.Following a review of the manufacturing process, their analysis confirmed that the separation is most likely to have been caused by an insufficient amount of solvent having been applied to the base of the burette during the assembly process.This is a manually performed assembly step and is likely to have occurred due to human error.A review of the production records for lot 21109297 did not identify any in-process testing failures or quality deviations which may have resulted in a report of this nature.A review of the customer feedback database indicates that this is a rare occurrence with a small number of similar reports against the 10037032 set in the past 12 months.
 
Event or Problem Description
It was reported that 5 bd alaris¿ smartsite¿ gravity burette sets leaked fluid during use.The following information was provided by the initial reporter: "customer has confirmed that the burette is leaking fluid during the use.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD ALARIS¿ SMARTSITE¿ GRAVITY BURETTE SET
Common Device Name
INTRAVASCULAR ADMINISTRATION SET
Manufacturer (Section D)
SISTEMAS MEDICOS ALARIS, S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer (Section G)
SISTEMAS MEDICOS ALARIS, S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key13759433
Report Number9616066-2022-00226
Device Sequence Number20283929
Product Code FPA
UDI-Device Identifier50885403244054
UDI-Public50885403244054
Combination Product (Y/N)N
Initial Reporter CountryNZ
PMA/510(K) Number
K780130
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date10/04/2024
Device Model Number10037032
Device Catalogue Number10037032
Device Lot Number21109297
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/23/2022
Supplement Date Received by Manufacturer03/16/2022
Initial Report FDA Received Date03/14/2022
Supplement Report FDA Received Date04/01/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/04/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1