• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DAVIS & GECK CARIBE LTD SURGIPRO II; SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLEN

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DAVIS & GECK CARIBE LTD SURGIPRO II; SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLEN Back to Search Results
Model Number VP713X-2
Medical Device Problem Code Break (1069)
Health Effect - Clinical Codes Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/25/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during a mitral valve repair with coronary artery bypass grafting, while suturing the anastomosis of aorta, the first thread broke after seventh throws while the second thread broke after three throws.The surgeon opened a new suture to complete the case.
 
Additional Manufacturer Narrative
Additional information: b5, g3, h6 medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during a mitral valve repair with coronary artery bypass grafting, while suturing the anastomosis of aorta, the first thread broke after seventh throws while the second thread broke after three throws.The surgeon opened a new suture to complete the case.There was no patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SURGIPRO II
Common Device Name
SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLEN
Manufacturer (Section D)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DR  0101
Manufacturer (Section G)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DR   0101
Manufacturer Contact
tracy landers
5920 longbow drive
boulder, CO 80301
3035816943
MDR Report Key13759795
Report Number9612501-2022-00408
Device Sequence Number20290157
Product Code GAW
UDI-Device Identifier10884521758216
UDI-Public10884521758216
Combination Product (Y/N)N
Initial Reporter StateIA
Initial Reporter CountryUS
PMA/510(K) Number
K010909
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2026
Device Model NumberVP713X-2
Device Catalogue NumberVP713X-2
Device Lot NumberD1E2903Y
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/25/2022
Supplement Date Received by Manufacturer03/21/2022
Initial Report FDA Received Date03/14/2022
Supplement Report FDA Received Date04/05/2022
Date Device Manufactured05/27/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age60 YR
Patient SexFemale
Patient Weight62 KG
-
-