• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHYSIO-CONTROL, INC. - 3015876 LIFEPAK® 15 DEFIBRILLATOR/MONITOR; AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHYSIO-CONTROL, INC. - 3015876 LIFEPAK® 15 DEFIBRILLATOR/MONITOR; AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE) Back to Search Results
Model Number 15
Medical Device Problem Code Failure to Deliver Shock/Stimulation (1133)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/14/2022
Type of Reportable Event Malfunction
Event or Problem Description
The customer contacted stryker to report that their device failed to shock.In this state the device may not be able to deliver defibrillation therapy if needed.Customer proceeded with another device.This issue is patient related; however there was no adverse event reported.
 
Additional Manufacturer Narrative
Stryker performed an initial evaluation of the customer's device and was unable to verify or duplicate the reported issue.Stryker continues to investigate the reported failure and will submit a supplemental report on this event to the fda as provided by 21 cfr 803.56.Stryker contacted the customer to request additional information on the patient.No response has been received from the customer.
 
Event or Problem Description
The customer contacted stryker to report that their device failed to shock.In this state the device may not be able to deliver defibrillation therapy if needed.Customer proceeded with another device.This issue is patient related; however there was no adverse event reported.
 
Additional Manufacturer Narrative
Stryker evaluated the software logs and verified the issue.After observing proper device operation through functional and performance testing, the device was returned to the customer for use.The cause of the reported issue could not be determined.
 
Additional Manufacturer Narrative
Section h10/h11 additional mfg narrative of supplemental medwatch report 001 indicates: stryker evaluated the software logs and verified the issue.Section h10/h11 additional mfg narrative of supplemental medwatch report 001 should indicate: the patient record files were reviewed and no device malfunction was found on the date of the incident.No parts were replaced.Results code has been updated from operational problem identified to no device problem found.
 
Event or Problem Description
The customer contacted stryker to report that their device failed to shock.In this state the device may not be able to deliver defibrillation therapy if needed.Customer proceeded with another device.This issue is patient related; however there was no adverse event reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEPAK® 15 DEFIBRILLATOR/MONITOR
Common Device Name
AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)
Manufacturer (Section D)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer (Section G)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer Contact
todd bandy
11811 willows road ne
redmond, WA 98052
4258674000
MDR Report Key13760030
Report Number0003015876-2022-00356
Device Sequence Number14936585
Product Code MKJ
UDI-Device Identifier00883873911778
UDI-Public00883873911778
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K142430
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 05/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number15
Device Catalogue Number99577-001242
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/16/2022
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 02/14/2022
Supplement Date Received by Manufacturer04/14/2022
05/13/2022
Initial Report FDA Received Date03/14/2022
Supplement Report FDA Received Date04/19/2022
05/18/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/23/2014
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient Age44 YR
Patient SexMale
Patient Weight132 KG
Patient EthnicityHispanic
Patient RaceWhite
-
-