MAUDE Adverse Event Report: DATASCOPE CORP. - FAIRFIELD MEGA 7.5FR. 40CC IAB; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE CORP. - FAIRFIELD MEGA 7.5FR. 40CC IAB; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0684-00-0295-02
Medical Device Problem Codes Leak/Splash (1354); Material Deformation (2976); Pressure Problem (3012)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/21/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that after approximately 10 days of intra-aortic balloon (iab) therapy the console generated an alarm after the patient was bathed.The insertion was reported to be right axillary, which is not the method described in the device instructions for use.The console had been generating a high pressure alarm with patient movement for several days.It was also noted that there was a kink in the iab.The console was now generating a low arterial pressure alarm.Blood was seen in the tubing.The console was turned off and the line was clamped and disconnected from the console.Iab removal was delayed due to act 199; act 160 at time of removal.The iab was removed about two hours later and a new one was placed in the left axillary.There was no patient harm or adverse event reported.
 
Additional Manufacturer Narrative
The device has been returned to the factory and is being evaluated.A supplemental report will be submitted when the evaluation is completed.Complaint record id # (b)(4).
 
Additional Manufacturer Narrative
The product was returned with the membrane completely unfolded with blood on the interior and exterior of the catheter.The extender tubing was also returned.A visual examination of the product detected the inner lumen within the catheter tubing near the y-fitting was completely separated within a kink approximately 76.2cm from iab tip.Additionally a second kink was found on the catheter tubing approximately 52.8cm from the iab tip.The inner lumen was found occluded with dried blood.The occlusion was unable to be cleared.An underwater leak test of the balloon, catheter, y-fitting, extracorporeal and extender tubing was performed and no other leaks were detected.The break found in the inner lumen appears to have been the result of a severe kink, which eventually failed causing reported problems the insertion was reported to be axillary, which is not the method described in the device instructions for use.This may have contributed to the iab failure.The evaluation confirmed the reported problem.A lot history record review was completed for the reported product.No nonconformances were found that are considered to be related to the event.Reference complaint #(b)(4).
 
Event or Problem Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEGA 7.5FR. 40CC IAB
Common Device Name
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - FAIRFIELD
15 law drive
fairfield NJ
Manufacturer (Section G)
DATASCOPE CORP. - FAIRFIELD
15 law drive
fairfield NJ
Manufacturer Contact
brian schaeffer
15 law drive
fairfield, NJ 
MDR Report Key13761545
Report Number2248146-2022-00211
Device Sequence Number14073879
Product Code DSP
UDI-Device Identifier10607567107981
UDI-Public10607567107981
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K120868
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/12/2024
Device Model Number0684-00-0295-02
Device Catalogue Number0684-00-0293
Device Lot Number3000183048
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/04/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/23/2022
Supplement Date Received by Manufacturer11/11/2022
Initial Report FDA Received Date03/14/2022
Supplement Report FDA Received Date11/11/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/13/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
AUTO CAT2 WAVE.
Patient Age40 YR
Patient SexFemale
Patient Weight90 KG
-
-