| Model Number |
SRLP-540 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Healthcare professional reported "right side rupture." device explanted.Right side.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation:.Visual analysis of the returned device identified: underweight, opening, crease fold and wear abrasion.A microscopic analysis was performed which identify crease sharp in the radius side with non-penetrating nicks around the opening.Based on the device analysis the final assessment is: -a crease sharp opening on radius due to a fold flaw opening.-non-penetrating nicks.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported "right side rupture." device has been explanted.
|
| |
|