| Model Number |
386803 |
| Medical Device Problem Code |
Material Protrusion/Extrusion (2979)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/27/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported while using bd cathena safety iv catheter the safety mechanism did not activate.There was no report of patient impact.The following information was provided by the initial reporter: the catheter did not secure itself at the time of removal, no accident to report.
|
| |
|
Additional Manufacturer Narrative
|
|
A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported while using bd cathena safety iv catheter the safety mechanism did not activate.There was no report of patient impact.The following information was provided by the initial reporter: the catheter did not secure itself at the time of removal, no accident to report.
|
| |
|
Additional Manufacturer Narrative
|
|
H6: investigation summary: one photo was received by our quality team for evaluation.From the photo, the safety mechanism was observed to not be activated.A device history record review found no non-conformances associated with this issue during production of this batch.As the actual sample was not returned, the root cause could not be established.
|
| |
|