|
Investigation ¿ evaluation: (b)(6) hospital informed cook of an event with an 85-year-old male patient with a type 1a endoleak.The complaint device was zenith flex aaa endovascular graft main body rpn: tffb-26-111-zt (product lot 2834215).The patient now presents with abdominal pain and an expanding aneurysm sac.It is not known if the patient had any pre-existing conditions or comorbidities, if the patient was on any medication regimen after original placement of the grafts, or if the patient was compliant with follow up protocol stated in the ifu.The patient needed a fenestrated cuff/branch graft to extend the proximal seal and has opted to be treated by open surgery.It was reported everything went fine during and post-surgery no complications.A review of the complaint history, device history record, drawing, instructions for use (ifu), manufacturing instructions, quality control, and specifications of the device were conducted during the investigation.The complaint device was not returned to cook for investigation.However, imaging was provided and reviewed by an expert reviewer.Findings included a focal limited type 1a endoleak adjacent to the proximal flex graft in the neck and at the proximal aaa sac.This lack of proximal seal is due to dilation of the proximal neck.The proximal graft diameter is 26 mm while the aortic neck diameter is 31 mm.It cannot be determined if this is due to disease progression or if the main body graft was undersized at the time of repair.Given the long postop interval and limited appearance of the type 1a endoleak, the likely scenario would be more recent disease progression.Additionally, a document based investigation evaluation was performed.After review of manufacturing documents cook has concluded that appropriate inspections are in place relative to the reported device failure.Cook reviewed the device history record for the final product and the subassembly work orders.Final product lot 2834215, the graft subassembly lot, and graft kit lot had no nonconformances reported.The graft subassembly lot was sewn by an approved vendor.Upon incoming inspection zero nonconformances were detected.There are no other complaints reported on the complaint lot.Therefore, cook concludes that the complaint device was manufactured per specification.There is no evidence of nonconforming devices from the complaint lot in house or in the field.The device was packaged with ifu t_zaaaf_rev5.The ifu includes the following, warnings, precautions, and instructions for proper placement of the device: 4 warnings and precautions: 4.1 general: ¿ additional endovascular interventions or conversion to standard open surgical repair following initial endovascular repair should be considered for patients experiencing an enlarging aneurysm, unacceptable decrease in fixation length (vessel and component overlap) and/or endoleak.An increase in aneurysm size and/or persistent endoleak or migration may lead to aneurysm rupture.¿ patients experiencing reduced blood flow through the graft limb and/or leaks may be required to undergo secondary interventions or surgical procedures.4.2 patient selection, treatment and follow-up: ¿ key anatomical elements that may affect successful exclusion of the aneurysm include severe proximal neck angulation (>60 degrees for infrarenal neck to axis of aaa or >45 degrees for suprarenal neck relative to the immediate infrarenal neck); short proximal aortic neck (<15 mm); an inverted funnel shape.(greater than 10% increase in diameter over 15 mm of proximal aortic neck length); and circumferential thrombus and/or calcification at the arterial implantation sites, specifically the proximal aortic neck and distal iliac artery interface.In the presence of anatomical limitations, a longer neck may be required to obtain adequate sealing and fixation.Irregular calcification and/or plaque may compromise the attachment and sealing at the fixation sites.Neck exhibiting these key anatomical elements may be more conducive to graft migration or endoleak.¿ multiple large, patent lumbar arteries, mural thrombus and a patent inferior mesenteric artery may all predispose a patient to type ii endoleaks.Patients with uncorrectable coagulopathy may also have an increased risk of type ii endoleak or bleeding complications.4.3 pre-procedure measurement techniques and imaging: ¿ clinical experience indicates that contrast-enhanced spiral computed tomographic angiography (cta) with 3-d reconstruction is the strongly recommended imaging modality to accurately assess patient anatomy prior to treatment with the zenith flex aaa endovascular graft.If contrast-enhanced spiral cta with 3- d reconstruction is not available, the patient should be referred to a facility with these capabilities.Lengths: ¿ the long-term performance of endovascular grafts has not yet been established.All patients should be advised that endovascular treatment requires life-long, regular follow-up to assess their health and the performance of their endovascular graft.Patients with specific clinical findings (e.G., endoleaks, enlarging.Aneurysm or changes in the structure or position of the endovascular graft) should receive enhanced follow-up.Specific follow-up guidelines are described in section 12, imaging guidelines and postoperative follow-up.¿ after endovascular graft placement, patients should be regularly monitored for perigraft flow, aneurysm growth or changes in the structure or position of the endovascular graft.At a minimum, annual imaging is required, including: 1) abdominal radiographs to examine device integrity (separation between components, stent fracture or barb separation) and 2) contrast and non-contrast ct to examine aneurysm changes, perigraft flow, patency, tortuosity, and progressive disease.If renal complications or other factors preclude the use of image contrast medica, abdominal radiographs and duplex ultrasound may provide similar information.4.4 device selection: ¿ strict adherence to the zenith flex aaa endovascular graft ifu sizing guide is strongly recommended when selecting the appropriate device size (tables 10.5.1 through 10.5.2).Appropriate device oversizing has been incorporated into the ifu sizing guides.Sizing outside of this range can result in endoleak, fracture, migration, device infolding or compression.4.5 implant procedure: ¿ inaccurate placement and/or incomplete sealing of the zenith flex aaa endovascular graft within the vessel may result in increased risk of endoleak, migration or inadvertent occlusion of the renal or internal iliac arteries.Renal artery patency must be maintained to prevent/reduce the risk of renal failure and subsequent complications.¿ fluoroscopy should be used during introduction and deployment to confirm proper operation of the delivery system components, proper placement of the graft, and desired procedural outcome.5 adverse events: 5.2 potential adverse events: ¿ aneurysm enlargement.¿ aneurysm rupture and death.¿ claudication (e.G., buttock, lower limb).¿ endoleak.¿ endoprosthesis: improper component placement; incomplete component.Deployment; component migration; suture break; occlusion; infection; stent fracture; graft material wear; dilation; erosion; puncture; perigraft flow; barb separation and corrosion.7 patient selection and treatment.7.1 individualization of treatment.¿ co-morbidities (e.G., cardiac, pulmonary, or renal insufficiency prior to surgery, morbid obesity).¿ patient¿s anatomical suitability for endovascular repair.¿ ability to tolerate general, regional, or local anesthesia.¿ iliofemoral access vessel size and morphology (minimal thrombus, calcium and/or tortuosity) should be compatible with vascular access techniques and accessories of the delivery profile of a 16 french to 22 french vascular introducer sheath.8 patient counseling information: ¿ the long-term performance of endovascular grafts has not yet been established.All patients should be advised that endovascular treatment requires lifelong, regular follow-up to assess their health and the performance of their endovascular graft.Patients with specific clinical findings (e.G., endoleaks, enlarging aneurysms or changes in the structure or position of the endovascular graft) should receive enhanced follow-up.Specific follow-up guidelines are described in section 12, imaging guidelines and postoperative follow-up.¿ patients should be counseled on the importance of adhering to the follow-up schedule, both during the first year and at yearly intervals thereafter.Patients should be told that regular and consistent follow-up is a critical part of ensuring the ongoing safety and effectiveness of endovascular treatment of aaas.At a minimum, annual imaging and adherence to routine postoperative follow-up requirements are required and should be considered a lifelong commitment to the patient¿s health and well-being.¿ physicians must advise each patient that it is important to seek prompt medical attention if they experience signs of limb occlusion, aneurysm enlargement or rupture.Signs of graft limb occlusion include pain in the hip(s) or leg(s) during walking or at rest or discoloration or coolness of the leg.Aneurysm rupture.May be asymptomatic, but usually presents as: pain; numbness; weakness in the legs; any back, chest, abdominal or groin pain; dizziness; fainting; rapid heartbeat or sudden weakness.Physicians should refer patients to the patient guide regarding risks occurring during or after implantation of the device.Procedure-related risks include cardiac, pulmonary, neurologic, bowel and bleeding complications.Device-related risks include occlusion, endoleak, aneurysm enlargement, fracture, potential for reintervention and open surgical conversion, rupture and death (see section 5.1, observed adverse event and section 5.2, potential adverse events).The physician should complete the patient i.D.Card and give it to the patient so that he/she can carry it with him/her at all times.The patient should refer to the card anytime he/she visits additional health practitioners, particularly for any additional diagnostic procedures (e.G., mri).12 imaging guidelines and postoperative follow-up: 12.1 general: ¿ the long-term performance of endovascular grafts has not yet been established.All patients should be advised that endovascular treatment requires lifelong, regular follow-up to assess their health and performance of their endovascular graft.Patients with specific clinical findings (e.G., endoleaks, enlarging aneurysms or changes in the structure or position of the endovascular graft) should receive additional follow-up.Patients should be counseled on the importance of adhering to the follow-up schedule, both during the first year and at yearly intervals thereafter.Patients should be told that regular and consistent follow-up is a critical part of ensuring the ongoing safety and effectiveness of endovascular treatment of aaas.¿ physicians should evaluate patients on an individual basis and prescribe their follow-up relative to the needs and circumstances of each individual patient.The recommended imaging schedule is presented in table 12.1.1.This schedule was used in the pivotal trial and is recommended even in the absence of clinical symptoms (e.G., pain, numbness, weakness).Patients with specific clinical findings (e.G., endoleaks, enlarging aneurysms or changes in the structure or position of the stent graft) should receive follow-up at more frequent intervals.¿ the combination of contrast and non-contrast ct imaging provides information on aneurysm diameter change, endoleak, patency, tortuosity, progressive disease, fixation length and other morphological changes.¿ duplex ultrasound imaging may provide information on aneurysm diameter change, endoleak, patency, tortuosity, and progressive disease.In this circumstance, a non-contrast ct should be performed to use in conjunction with the ultrasound.Ultrasound may be a less reliable and sensitive diagnostic method compared to ct.12.6 additional surveillance and treatment: ¿ aneurysms with type i endoleak.After review of the ifu, cook has concluded the device labeling contains appropriate warnings, precautions and instructions to the user related to the reported failure.Based on the information provided, imaging review and the results of the investigation, a definitive cause could not be established.While the cause in this case is not known, it should be noted that endoleak is a known inherent risk of the device.The appropriate personnel have been notified.Per the risk assessment, no further action is required.Cook will continue to monitor for similar complaints.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|