U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORPORATION ENDOSCOPE 5.5MM; LAPAROSCOPE, GENERAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORPORATION ENDOSCOPE 5.5MM; LAPAROSCOPE, GENERAL Back to Search Results
Model Number MCENDO550
Medical Device Problem Code Optical Problem (3001)
Health Effect - Clinical Code Insufficient Information (4580)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Terumo has received the device for evaluation; however, the investigation has yet to be completed.Terumo plans on submitting a follow-up report when the investigation is complete and when more information becomes available.(b)(4).
 
Event or Problem Description
The user facility reported to terumo cardiovascular an obstructed view.It is still unknown when the event occurred, if it resulted in a delay in the procedure, if the product was changed or if the surgery was completed successfully or if it caused or contributed to an injury to the patient or if there was a blood loss.Terumo continues an attempt to gain more information regarding this event from the user facility.
 
Additional Manufacturer Narrative
This follow-up report is submitted to fda in accord with applicable regulations ¿ and as indicated by terumo cardiovascular systems in the initial report submitted to the fda on march 14, 2022.Upon further investigation of the reported event, the following information is new and/or changed: d4 (additional device information - added exp date); g3 (date received by manufacturer) ; g6 (indication that this is a follow-up report) h2 (follow-up due to additional information and device evaluation); h3 (device evaluated by manufacturer); h4 (device manufacture date); h6 (identification of evaluation codes 10, 3331, 3224, 19).Type of investigation #1: 10 - testing of actual/suspected device; type of investigation #2: 3331 - analysis of production records; investigation findings: 3224 - optical problem identified; investigation conclusions: 19 - cause traced to user.Upon evaluation of the device the complaint was confirmed.The complaint sample was returned and visually inspected.A monocular was used to inspect the device and it was discovered that there was a crack on one of the internal lenses which resulted in a blurry visual field.Although a definitive root cause could not be determined, it is likely that the crack in the lens was due to improper handling of the device.All available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOSCOPE 5.5MM
Common Device Name
LAPAROSCOPE, GENERAL
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
125 blue ball road
elkton MD 21921
Manufacturer (Section G)
SAME
Manufacturer Contact
jamie quinlan
125 blue ball road
elkton, MD 21921
8002837866
MDR Report Key13763342
Report Number1124841-2022-00043
Device Sequence Number2441971
Product Code GCJ
UDI-Device Identifier00699753018952
UDI-Public(01)00699753018952
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K092789
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberMCENDO550
Device Catalogue NumberN/A
Device Lot Number800508
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/25/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/17/2022
Supplement Date Received by Manufacturer04/28/2022
Initial Report FDA Received Date03/14/2022
Supplement Report FDA Received Date04/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/17/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-