MEDELA LLC SWING MAXI; PUMP, BREAST, POWERED Back to Search Results
Model Number 101043614
Date of Event 02/13/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
On 02/24/2022, the customer called and spoke with customer service and stated she didn't want to use the swing maxi pump and didn't want to troubleshoot.She would like a pump in style as her replacement pump.The customer was sent a replacement pump and return of her original pump was requested for testing/evaluation.In follow up with a complaint handler on 02/24/2022, the customer indicated that she developed clogged ducts and milk blisters and was prescribed medication.She stated her clogged ducts are reoccurring and she is being diligent to express as often and long but is using her previous pump in style breast pump.The device was returned with the customer's parts and accessories and was evaluated on 03/11/2022.The device passed suction and cycle specifications.Medela is filing this report, which is considered a serious injury as it required medical attention (medication was prescribed).
 
Event or Problem Description
On (b)(6) 2022, the customer emailed medela llc and alleged that her recently purchased swing maxi breast pump she used for a full day and night was painful and caused clogged ducts.She went to the hospital due to her breast being hard.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SWING MAXI
Common Device Name
PUMP, BREAST, POWERED
Manufacturer (Section D)
MEDELA LLC
1101 corporate dr
mchenry IL 60050
Manufacturer (Section G)
MEDELA LLC
1101 corporate dr
mchenry IL 60050
Manufacturer Contact
dave kurudza
1101 corporate drive
mchenry, IL 60050
8004358316
MDR Report Key13763723
Report Number1419937-2022-00026
Device Sequence Number5033030
Product Code HGX
UDI-Device Identifier00020451436142
UDI-Public020451436142
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
K210759
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 03/14/2022
1 Device was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model Number101043614
Device Catalogue Number101043614
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/07/2022
Distributor Facility Aware Date02/24/2022
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/24/2022
Initial Report FDA Received Date03/14/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/16/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
-
-