PHILIPS NORTH AMERICA LLC HEARTSTART MRX -EMS DEFIBRILLATOR Back to Search Results
Model Number M3536A
Device Problem Sparking (2595)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/09/2022
Event Type  malfunction  
Manufacturer Narrative
Reporting institution phone: (b)(6).
 
Event Description
It was reported to philips that their was sparks during shock delivery.
 
Manufacturer Narrative
The field service engineer (fse) evaluated the device and found the paddle trays need to be replaced.Upon conclusion of the evaluation, it was determined that the paddle trays require replacement.They were replaced.The device passed testing and was put back into service.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEARTSTART MRX -EMS DEFIBRILLATOR
Type of Device
DEFIBRILLATOR
Manufacturer (Section D)
PHILIPS NORTH AMERICA LLC
22100 bothell everett highway
bothell WA 98021
Manufacturer (Section G)
PHILIPS NORTH AMERICA LLC
22100 bothell everett highway
bothell WA 98021
Manufacturer Contact
dana tackett
22100 bothell everett highway
bothell, WA 98021
9095703538
MDR Report Key13764562
MDR Text Key287180324
Report Number3030677-2022-01554
Device Sequence Number1
Product Code MKJ
UDI-Device Identifier00884838006652
UDI-Public00884838006652
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
K031187
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberM3536A
Device Catalogue NumberM3536A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received Not provided
Initial Date FDA Received03/14/2022
Supplement Dates Manufacturer Received03/14/2022
Supplement Dates FDA Received05/19/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-