• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION INTERLOCK-35; DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION INTERLOCK-35; DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION Back to Search Results
Model Number 83786
Medical Device Problem Code Separation Failure (2547)
Date of Event 01/05/2022
Type of Reportable Event Malfunction
Event or Problem Description
Reportable based on device analysis completed on 21feb2022.It was reported that the coil could not be deployed.A 8mm x 20cm interlock coil was selected for use.During the procedure, it was noted that the coil could not be deployed normally.The procedure was completed with a different device.There were no patient complications reported and the patient was stable.However, device analysis revealed that there were missing fiber bundles.
 
Additional Manufacturer Narrative
(b)(6).Device evaluated by manufacturer: the device was returned for analysis.The main coil, the delivery wire and introducer sheath were returned for this complaint.The main coil and delivery wire were interlocked within the introducer sheath.The delivery wire and introducer sheath were inspected, and no damages were observed.The main coil was found kinked and stretched.The delivery wire was advanced through the introducer sheath, resistance was noticed, it was necessary to pull it backwards to release the main coil.The zap tip has a smooth surface.The interlocking arm was inspected, and no damages were found.The dimensional inspection revealed that the outer diameter (od) at the zap tip and primary coil were found to be within specification.Also, proximal fiber bundles were found to be within specification, however, distal fiber bundles were not within the specifications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTERLOCK-35
Common Device Name
DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORK LIMITED
model farm road
cork
EI  
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key13764688
Report Number2134265-2022-02662
Device Sequence Number4993362
Product Code KRD
UDI-Device Identifier08714729793045
UDI-Public08714729793045
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K133208
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 03/14/2022
1 Device was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/29/2024
Device Model Number83786
Device Catalogue Number83786
Device Lot Number0027058814
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/03/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/21/2022
Initial Report FDA Received Date03/14/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/30/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
-
-