MAUDE Adverse Event Report: ZOLL CIRCULATION ZOLL IVTM QUATTRO CATHETER; CENTRAL VENOUS CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZOLL CIRCULATION ZOLL IVTM QUATTRO CATHETER; CENTRAL VENOUS CATHETER Back to Search Results
Model Number IC-4593
Medical Device Problem Code Use of Device Problem (1670)
Health Effect - Clinical Code Hemorrhage/Blood Loss/Bleeding (1888)
Date of Event 02/17/2022
Type of Reportable Event Malfunction
Event or Problem Description
After 40 hours of ivtm therapy for a patient with cardiopulmonary arrest, the therapy was paused and the start-up kit (suk) (lot #158572) was disconnected from the catheter for the patient examination and connected back 30 minutes later to continue the therapy.Per the reporter, the pinwheel was spinning after the catheter was reconnected to the suk and no kinks or obstructions to the suk tubing were noted.The customer observed bleeding from the catheter insertion site.The pressure was applied to mitigate the bleeding.Twenty minutes after the therapy restarted the user heard a bursting noise and the lfl tube portion of the suk just after the rotor pump was observed swollen and torn with saline leaking externally.The thermogard console generated the air trap alarm and the therapy was stopped.The customer verified that the catheter was secure and in the correct depth position.The customer suspected the catheter leak.The catheter was removed from the patient, however, after performing a balloon leak test on the catheter, the saline leak was not confirmed.The treatment was continued using the alternative temperature management device (blanket roll).The customer did not provide the reason for not continuing the ivtm therapy, however, the console was confirmed to be functional.No consequences or impact to the patient.In addition, the customer reported the bubbles in the suk and water droplets formation inside the lid of the roller pump after the second day of the treatment.
 
Additional Manufacturer Narrative
The zoll icy catheter was not returned to zoll for evaluation.A supplemental report will be filed if and when the product is returned and investigation has been completed.Event of bleeding was not serious, the patient did not require treatment, and event resolved without sequelae.Event was probably related to the zoll catheter due to relevant timing and location at insertion site.Bleeding at insertion site is anticipated event and could potentially occur with usage of any catheter.
 
Additional Manufacturer Narrative
The functional testing of the returned quattro catheter (lot # 155959) was performed and there was no device malfunction observed.The catheter worked as intended.A visual examination of the returned catheter was performed and found no physical damage to the catheter.A functional pressure leak test of the returned catheter was performed.All infusion ports and extension tubes were flushed without resistance.The catheter was connected to a pressurized inflation device, and the balloons were fully inflated when pressurized up to 100 psi, and no issues were observed.The balloons did not leak during inflation and deflation.Event of bleeding was not serious, the patient did not require treatment, and event resolved without sequelae.Event was probably related to the zoll catheter due to relevant timing and location at insertion site.Bleeding at insertion site is anticipated event and could potentially occur with usage of any catheter.The event is probably related to the device and is not related to the procedure.Additional information, correction, b5: (describe event or problem) was updated based on the received quattro catheter lot # 155959.
 
Event or Problem Description
After 40 hours of ivtm therapy for a patient with cardiopulmonary arrest, the therapy was paused and the start-up kit (suk) (lot # 158572) was disconnected from the quattro catheter (lot # 155959) for the patient examination and connected back 30 minutes later to continue the therapy.Per the reporter, the pinwheel was spinning after the catheter was reconnected to the suk and no kinks or obstructions to the suk tubing were noted.The customer observed bleeding from the catheter insertion site.The pressure was applied to mitigate the bleeding.Twenty minutes after the therapy restarted the user heard a bursting noise and the lfl tube portion of the suk just after the rotor pump was observed swollen and torn with saline leaking externally.The thermogard console generated the air trap alarm and the therapy was stopped.The customer verified that the catheter was secure and in the correct depth position.The customer suspected the catheter leak.The catheter was removed from the patient, however, after performing a balloon leak test on the catheter, the saline leak was not confirmed.The treatment was continued using the alternative temperature management device (blanket roll).The customer did not provide the reason for not continuing the ivtm therapy, however, the console was confirmed to be functional.In addition, the customer reported the bubbles in the suk and water droplets formation inside the lid of the roller pump after the second day of the treatment.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZOLL IVTM QUATTRO CATHETER
Common Device Name
CENTRAL VENOUS CATHETER
Manufacturer (Section D)
ZOLL CIRCULATION
2000 ringwood ave.
san jose CA 95131
Manufacturer (Section G)
ZOLL CIRCULATION
2000 ringwood ave.
san jose CA 95131
Manufacturer Contact
kim thoa nguyen
2000 ringwood ave.
san jose, CA 95131
4084192922
MDR Report Key13764708
Report Number3010617000-2022-00266
Device Sequence Number20082021
Product Code NCX
UDI-Device Identifier00849111075527
UDI-Public00849111075527
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K101987
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberIC-4593
Device Catalogue Number8700-0783-03
Device Lot Number155959
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/17/2022
Supplement Date Received by Manufacturer04/14/2022
Initial Report FDA Received Date03/14/2022
Supplement Report FDA Received Date05/09/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/01/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-