Medtronic received information regarding a navigation system being used outside of a procedure.It was reported that after changing the input / output (i/o) panel during a different case on the site, a system checkout was performed and it was noted that the site's camera was connected and initially green in the software.When a demo case was done, it said localizer not connected.Both leds (light-emitting diode) on the camera were green.The clinical consultant (cc) troubleshot by re-seating the camera cable from the sensor control unit (scu) and attempted bypassing the input / output hub cable from the central processing unit (cpu) directly to the scu but no change in software.After checking the network device interface (ndi) toolbox utility, the cc found no faults listed on camera or scu and no errors in the position sensor unit (psu) logs.However multiple internal strobe errors were noted in the scu logs.The probable cause was likely a secondary issue not related to the input/output (i/o) panel exchange.There was no patient present when the event occurred.
|
|
Concomitant medical products: other relevant device(s) are: product id: 9735347r.No parts have been received by the manufacturer for evaluation.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|