MAUDE Adverse Event Report: CIRPRO DE DELICIAS S.A. DE C.V. NPWT ALLY; NEGATIVE PRESSURE WOUND THERAPY POWERED SUCTION PUMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CIRPRO DE DELICIAS S.A. DE C.V. NPWT ALLY; NEGATIVE PRESSURE WOUND THERAPY POWERED SUCTION PUMP Back to Search Results
Model Number 68-8888
Medical Device Problem Codes No Audible Alarm (1019); Suction Failure (4039)
Date of Event 02/16/2022
Type of Reportable Event Malfunction
Event or Problem Description
Customer report indicated that the motor on the device was not running when the canister was removed, however the unit still had plenty of therapy time remaining.The unit tech was unsure if the dressing was sealed and depressed while on the patient.The device was tested once it was returned to the unit¿s department.Testing confirmed that there was no suction.Unit tech was unsure if the device sound setting was changed when it was used but confirmed that the device did not alarm when they tested the unit.There was no injury to patient and patient is doing fine and is still using the backup device provided by the facility.
 
Additional Manufacturer Narrative
The device with lot number c01201904740 was manufactured in 2019.The device history record of the complaint lot was reviewed.The lot was inspected and released in compliance with all requirements.No sample was provided by the customer; therefore, an evaluation of the complaint device for deficiency of construction could not be performed.Based on the investigation, no specific assignable cause can be determined as there is limited information.Without a specific assignable cause, no action will be taken at this time.We will continue to monitor trends based on this type of issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NPWT ALLY
Common Device Name
NEGATIVE PRESSURE WOUND THERAPY POWERED SUCTION PUMP
Manufacturer (Section D)
CIRPRO DE DELICIAS S.A. DE C.V.
las virgenes industrial park,
delicias, chihuahua 33019
MX  33019
Manufacturer (Section G)
CIRPRO DE DELICIAS S.A. DE C.V.
las virgenes industrial park,
delicias, chihuahua 33019
MX   33019
Manufacturer Contact
patricia tucker
3651 birchwood dr.
waukegan, IL 60085
8478874151
MDR Report Key13764911
Report Number1423537-2022-00734
Device Sequence Number5008102
Product Code OMP
UDI-Device Identifier10885380153204
UDI-Public10885380153204
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 03/14/2022
1 Device was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number68-8888
Device Catalogue Number68-8888
Device Lot NumberC01201904740
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/17/2022
Initial Report FDA Received Date03/15/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
-
-