C.R. BARD, INC. (BASD) -3006260740 POWERPICC 5F DUAL-LUMEN POLYURETHANE CATHETER RADIOLOGY BASIC TRAY; CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS Back to Search Results
Model Number N/A
Medical Device Problem Code Material Frayed (1262)
Date of Event 12/09/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The following were reviewed as part of this investigation: patient severity, sample analysis, and applicable fmea documents.Based on a review of this information, the following was concluded: the complaint of complications during the insertion process was confirmed.One unraveled guidewire was received within an introducer needle.A break was observed in the coiled end of the guidewire, which allowed the coil wire to stretch and unravel over the core wire.The outside diameter of the guidewire was within specification.The inner edge of the needle bevel exhibited deformation, which is consistent with the guidewire being retracted through the needle.The instructions for use (ifu) caution, ¿if the guidewire must be withdrawn while the needle is inserted, remove both the needle and wire as a unit to prevent the needle from damaging or shearing the guidewire.¿ no damage associated with the manufacturing process was observed on the returned sample.A lot history review (lhr) review is not possible, as no manufacturing lot number has been provided by the complainant.
 
Event or Problem Description
It was reported that during catheter insertion, the catheter was found difficult to insert.After removal, the wire and catheter were found deformed.(b)(6) 2022- the returned guidewire was broken.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERPICC 5F DUAL-LUMEN POLYURETHANE CATHETER RADIOLOGY BASIC TRAY
Common Device Name
CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
kelsey erickson
605 north 5600 west
salt lake city 84116
8015950700
MDR Report Key13764981
Report Number3006260740-2022-00793
Device Sequence Number17409052
Product Code LJS
UDI-Device Identifier00801741027932
UDI-Public(01)00801741027932
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
K051672
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 02/26/2022
1 Device was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number3275155
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/05/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/25/2022
Initial Report FDA Received Date03/15/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
-
-